9.1 Pause mode - reasons
The following events can lead to pause mode
1
:
Manual pausing by user (pressing the
„pause“ button)
Opening the door (see item 9.2)
Thermal fuse failure
Air circulation malfunction
Turntable malfunction
Heating malfunction
Be aware
While the underlying cause of the malfunction has not been corrected, you cannot leave pause
mode.
9.2 Pausing when the door is opened
When the door is opened, the light will automatically come on and the extractor (exhaust fan) will be
switched on for 60 seconds
2
, providing that the extractor is set up as
„auto“.
Optional displays /
Extractor (exhaust fan)
Exhaust fan switch-on for 60 secunds , if the
nominal value is
set to “auto”.
Shortcut buttons/ Operating mode
9. Air circulation off
3
10. Steam flap off
11. Turntable
– off
4
12. Resume program
13. Menu (item 3)
14. Light off
15. Extractor (exhaust fan) off
16. Terminate all operations. Back to
readiness mode (see item 7)
Information bar
System status display (flashing)
In pause mode the reason for pausing the
program will also be shown.
10
Alarm signal
If an alarm incident has occurred, for example through digital inputs, limit value alarms or sensor failure,
instead of chamber symbol the alarm symbol will displayed.
On the status bar you can see the alarm and the cause
(e.g. chamber sensor break).
Press the alarm
symbol to acknowledge the alarm and to stop the horn. The message will remain on the status bar.
Alarm signal
1
If required, further events can be configured for pause mode
2
Configuration setting in the basic configuration: Code 4552/ Input 1 „Door“; 1004/ Relay 10, Par.1 = 60s
3
Manual switch-on possible. Manual activation is shown by the „M“ in the button.
4
Manual approach of the end position is possible.
1
2
3
4
5
6
7
8