8
Technical Data
Standard
Performance and Specifications
Input specifications
Maximum input amplitude:
±
10 V or less
Input connector: 10-pin connector
Isolation method
Analog signal transmission through hot coupler
Constant-voltage section
Rough adjustment: 0.1, 1, 10 V (3-step adjustment)
Fine adjustment: Control for continuous adjustment of all ranges
Magnification meter:
×
1 to
×
10 (23-step control)
Output maximum current: 10 mA
Output impedance: 200
Ω
or less (2 k
Ω
or less in the 1 V range)
Pulse response: 10
µ
s or less rising for all ranges
Excessive deviation: 5% or less for all ranges
Constant-current section
Rough adjustment: 10
µ
, 100
µ
, 1 m, 10 mA (4-step adjustment)
Fine adjustment: Control for continuous adjustment of all ranges
Pulse response: 10
µ
s or less (RL = 10 k
Ω
) rising for all ranges
Excessive deviation: 5% or less for all ranges
Output current characteristics
Guaranteed stimulation current range (see diagram below)
Output insulation: Between output terminals and cabinet 100 M
Ω
or more (500 DC V)
Stray capacitance: Between output terminals and cabinet 150 pF
or less
Display
Indication: 3-digit LED indi unit
Accuracy: 5% or less
Output polarity switch
Provided
Output OFF function
STIM ON/OFF switch Provided
OVERLOAD detection function
Provided
An alarm is generated when the guaranteed range of stimulation
current is exceeded.