Keypoints:
•
It can show four pairs of marker on time at the most, but there is only one pair or one single
marker which is in activated status each time.
•
You can use numeric keys, knobs or direction keys to modify this parameter. The detail
method refers to the specification in “Parameter Setting”.
1. Marker
Choose one of four markers, the default one is the first. After choosen, you can set the mode of
marker, trace that is marked, reading method and other parameters. The marker which is opened
currently should be marked on the trace chosen by marker trace. The reading on the marker from
which is in activated status currently is shown on the right corner of the screen in active function
area.
Marker Paremeter
Parameter
Instruction
Default
Center Frequency
Range
0 ~ 1.6GHz (1.6GHz)
Unit
Reading = Frequency (or period), unit is GHz, MHz, kHz, Hz (or ks, s, ms, us,
ns, ps)
Reading = Time (or reciprocal of time), unit is ks, s, ms, us, ns, ps (or GHz,
MHz, kHz, Hz)
Knob Step
Reading = Frequency (or period), Step = Sweeping width/(sweep point - 1),
Reading = Time (or reciprocal of time), Step = Sweeping time/(sweep point – 1)
Direction Key
Step
Reading = Frequency (or period), Sweeping width/10
Reading = Time (or reciprocal of time), Step = Sweeping time/10
2. Normal
One type of marker, used for measuring X value (frequency and time) and Y value (amplitude) of
one point on the trace. When choose “Normal”, there will be a marker marked by the current
marker on the trace, like “1”.
Keypoints:
•
If there is no active marker, then you need to activate a marker in center frequency of current
trace.
•
Move the marker through inputing value by numberic keys, knobs or direction keys. The
reading of current marker is shown on the right corner of screen.
•
The resolution of reading on X-axis (frequency or time) is related with sweep width. Reduce
the sweep width can get higher resolution.
3. Off
Close the marker chosen currently, and the marder information and relative function shown in the
screen will be closed too.
4. Marking trace
Choose the trace marking by current marker as 1, 2, 3, 4.
5. Readout
Set the reading method of marker X-axis, each marker can be set different reading type. This set
only change the method of reading, not actual value. This set will influence the marker reading in
active function area and the corner of screen.
a) Frequency
“Normality” marker shows absolute frequency when choose this type of reading method.
Summary of Contents for DSA2016TG
Page 1: ...USER S MANUAL Handheld Spectrum Analyzer DSA2016TG V1 0 0 ...
Page 7: ...Chapter1 Quick Start u Instrument Introduction u General Check u Preparation ...
Page 15: ...2 3 2 Left Panel Left Panel 5 Headphone Connects to a headphone 6 SD card slot 5 6 ...
Page 66: ...input 220VAC 15 AC frequency range 40Hz to 60Hz Power consumption Maximum 32W ...