ADE M400660 Operating Manual Download Page 9

M400660-210326-Rev005-UM-e 

 

Connecting additional devices 

Connection of printer, PC or remote-display will be done with the RS-232 interface. For connection of 
additional  devices  kindly  ask  your  local  representative  or  authorized  ADE  dealer  for  further 
instructions. 

 

Print out a record 

The  scales  must  be  at  standstill.  The  standstill  status  symbol 

  must  appear  at  the  screen.  Press 

the ENTER key to send the data to the serial interface.  

Note:  Printing  is  not  possible  if  the  net  weight  is  0.  The  print  function  only  produces  one  print-out.  A 
reset is required to print out another copy.  

 

Error Messages 

Error Codes 

Description 

Solution 

 

 

Load cell wires to indicator are 
incorrectly connected; or shorted, or 
opened; or ADC/ load cell are being 
damaged  

Make sure wires are in good order and 
condition and correctly connected. 
Replace load cell or ADC chip. Service 
is required.

 

 

 

Weight reading exceeds the zero limit 
after switching the scale on. 

Make sure scale platform is empty. 

 

Weight reading is below the zero limit 
after switching the scale on. 

Install platform on scale. Otherwise the 
zero calibration/ adjustment have to be 
done by authorized personnel. 

 

Weight reading exceeds load limit.  

Reduce load on scale until weight 
value can be displayed.  

 

Weight reading below load limit. 

Service by authorized personnel 
required. 

 

Configuration or Calibration 
parameters are not correctly set. 

Contact your local dealer / supplier for 
calibration purpose. 

 

Capacity parameters are not correct. 

Contact your local dealer / supplier for 
a new set-up. 

 

Calibration error. The incoming signal 
of loaded weight is too small, too big, 
unstable or un-linear. 

Contact your local dealer / supplier for 
a new set-up. 

Not turn on. 

Power cord is not properly 
connected. Power outlet not 
supplying electricity. Batterie is 
discharged. Other failure. 

Check power cord connections. Make 
sure power cord is plugged into the 
power outlet. Check power source. 
Replace batteries. 

Cannot zero the 

display or will not 

zero when turned 

on. 

Load on scale exceeds allowable 
limits, Load on scale is not stable or 
a load cell damage. 

Remove load from the scale; wait for 
load to become stable. Service 
required. 

No Battery symbol 

or display of Lo.bAt  

Batteries are discharged. 

Charge batteries. 

Summary of Contents for M400660

Page 1: ...Operating Manual International Range Electronic Chair Scales M400660 M400660 120516 Rev002 UM e M400660 210326 Rev005 UM de ...

Page 2: ...Key pad 5 Operation 6 Positioning of patient 6 Correct weighing 6 How to change the unit 6 Tare function 6 Weighing with preset tare values 7 HOLD function 7 BMI Body Mass Index 7 Evaluation of the Body Mass Index 7 Connecting additional devices 8 Print out a record 8 Error Messages 8 Cleaning and Maintenance 9 Cleaning 9 Storage and Transport 9 Warranty Liability 9 Technical data 10 CE marking an...

Page 3: ...about how to use the products or about this or other service manuals feel free to contact your local ADE representative or visit our web page at www ade hamburg de In order to achieve the best possible results kindly read and follow the operation manual in all respects Check carton contents Chair Scale assembled Mains adapter Operating manual Electronic Chair Scales Series M4006600 Configuration o...

Page 4: ...other interference Portable and mobile HF communication systems may influence electronic medical devices All actions and operations described in this operation manual may only be carried out by professional personnel authorized by ADE or official appointed dealers Assembly and set up The ADE chair scale is already been delivered in a fully assembled condition Please insure that your scale is being...

Page 5: ... battery cables carefully from the housing connect the wires and insert the battery into the housing 3 Close the battery compartment again Only use original rechargeable batteries and charger units from ADE The warranty does not cover any damage caused by using other types of batteries or charger units Display Front view and side view of the display Display description ...

Page 6: ...it is displayed in pounds kg Displayed when the weight is displayed in kilograms Key pad Name Description HOLD Activates the Hold Function PRINT Transfer of data to serial communication port RS 232 BMI After the user steps on the scale enter same in order to proceed with the BMI mode UNIT Switch button between kgs and lbs TARE PRESET Activates the tare function and the Preset Tare Function ZERO ON...

Page 7: ...et rest on the footrest The utmost accuracy has been achieved after the weight is being displayed and the symbol appears To turn off the scale press ON OFF for a period of minimum 3 seconds Using the scale on mains it may remain permanently switched on or can be shut off manually by pressing key How to change the unit The scale offers the possibility to select between the weighing mode kg and lb T...

Page 8: ...he patient into the chair Take care that the feet are being rested on the footrest Wait until the weight is being displayed in a stable condition and appears on the screen Press the BMI key shortly The scale starts automatically with the last input Use the arrow buttons to enter the height of the patient Confirm the height with the enter button The BMI is now being displayed To correct the given h...

Page 9: ...ter switching the scale on Install platform on scale Otherwise the zero calibration adjustment have to be done by authorized personnel Weight reading exceeds load limit Reduce load on scale until weight value can be displayed Weight reading below load limit Service by authorized personnel required Configuration or Calibration parameters are not correctly set Contact your local dealer supplier for ...

Page 10: ...delivery applying to defects occurred due to poor material or workmanship presentation of invoice is necessary All moveable parts batteries cables mains units rechargeable batteries etc are excluded Warranty does not include natural wear damages caused by incorrect or negligent handling overuse or unsuitable or improper storage The cost of transport will be borne by the customer should the equipme...

Page 11: ...dapter AC Adapter 6 9VDC 500mA Central negative 4xAA NIMH rechargeable batteries pack Display 6 digit 7 segment 1 25mm LCD with 16 annunciators and blue backlight Working temperature 10 C to 40 C Storage temperature 20 C to 70 C Humidity 10 to 90 RH without condensation Voltage 5Vdc Max Current 60mA can power 4 350 ohm bridge take care the limitation when USB is power supply Serial port Full duple...

Page 12: ...Germany 49 40 432 776 0 49 40 432 776 10 info ade germany de www ade germany de Declaration of conformity by the manufacturer These scales have been manufactured in accordance with the harmonised European standards They conform to the regulations of the EC directives listed below EMC Directive 2014 30 EC Low Voltage Directive 2014 35 EC RoHS Directive 2011 65 EC in the currently valid versions Thi...

Page 13: ...M400660 210326 Rev005 UM e 12 ...

Page 14: ...13 M400660 210326 Rev005 UM e ...

Page 15: ...M400660 210326 Rev005 UM e 14 ...

Page 16: ...15 M400660 210326 Rev005 UM e Dealer ...

Reviews: