8.4.3 Error messages
Error message
Description
Troubleshooting
[Lo]
Empty battery.
Insert new batteries.
Use the scale in mains
operation.
[uLoad]
Underload (-20d)
Switch the scale off and on
again.
[oLoad]
Overload (-9d)
Unload the scale. Zero range or
weighing capacity exceeded
[no 0.00]
No zero point available.
Zero the scale.
„-----“
Negative weight values (e.g. in
the tare function) are not
displayed.
Load the scale.
„-----“
Scale is loaded too slowly
during the HOLD function.
Unload the scale. Wait until
>0< is shown in the display.
Load the scale again.
8.5
Maintenance
To ensure correct measurement, maintenance and repair should only be carried out
by authorised personnel.
To prevent the intended level of accuracy the product must be set up carefully and services
regularly. We recommend having it serviced every 3 to 5 years depending on how often the scales
are used.
8.6
Storage and Transport Conditions
Keep all original packaging materials and components for eventual return of the scale to avoid
damage during transport; these are not covered by the warranty.
To avoid damage disconnect all cables before transport.