6. Cleaning
Cleaning can be done with lukewarm water and a suitable all-purpose cleaning agent. Use a cloth or soft
sponge to wash the product. Dry the product with a wrung cloth.
For surface disinfection, chlorine-based disinfectants may be used in accordance with the product
specification and safety precautions for these products, however, not at a solution greater than 1.000 ppm of
active chlorine.
Likewise, ethanol-based disinfectants may be used in accordance with the product specification and safety
precautions for these products.
The following agents have been tested by Ropox, and can be used, as long as they are diluted according to
the instructions on the container.
6.1 Preapproved disinfection products
Tradename
Solution
Diversey Suma Tab D4
1.000 ppm
Rodalon ®
Ethanol 70% v/v
6.2 Preapproved cleaning products [Toilet lifter]
Tradename
Solution
S90 Sanirens
1-3% peracid
For the Toilet lifter with a spiral cable going to its control, the following agents have been recommended by
its manufacturer for the spiral cable.
6.3 Preapproved cleaning products [Spiral cable]
Tradename
Solution
Sekumatic FDR or FRE
3mL/L [pH 11.5-12]
Neodisher Dekonta
Thermosept NDR
0.5% [pH 5-8]
Immidiately after cleaning or disinfection, the product must be wiped down with a cloth wrung in clean water,
then a dry rag, as some materials may otherwise become discolored and stained, and the surface coating
may deteriorate.
Keep in mind, after prolonged use of the product, with regular cleaning intervals, it can be expected that
certain surfaces will not have the same sheen as initially. The product can be disinfected regardless.
Note!
Do not use cleaning agents containing abrasive e.g., scouring powder, steel wool, scouring
sponge.
This product is not designed to be sterilized. Autoclaving and sterilization beyond normal
means may change the product safety and function.