28
Figure 13 PVAO zero
2. Main Variable Zero
Set the zero point of the pressure transmitter to correct the zero offset of the transmitter. Under the zeroing mode,
please select the pressure module connected to the device as the standard module. Then the upper channel in the
interface displays the pressure value output by the transmitter, and the lower channel displays the pressure value
measured by the selected standard module of the machine.
Click the apply button
on the upper right, the device will send a command to calibrate the transmitter.