39
4. Finish the task
After the task is finished, click the save button
to go to the Saving interface, fulfill the information required and
save the task data of the result.
Table 14 Save task data setting
Subject
Valid value
Description
Name
Numbers, letters, symbols
Task name
Operator
Numbers, letters, symbols
Information for operator
Date and time
Valid date and time
Date and time of the task performing
Note
Numbers, letters, symbols
Additional information
6.2 Pressure transmitter (voltage, current and HART)
1. Connect to DUT
•
Connect the internal pressure module. If using an external pressure module, connect them to LEMO port A
or B.
•
Connect the DUT output with the calibrator according to the signal type, current (Figure 7), voltage (Figure
6), or HART (Figures 10 and 11).
2. Create the task
Create the corresponding DUT tasks as described above, and enter the necessary information.
Table 15 Pressure transmitter task setting
Subject
Valid value
Description
Pressure type
Gauge / absolute / differential
Pressure type of DUT
Input range
Numbers
Input range of DUT