26
transmitter
Device Version
Read only
Device version of the
transmitter
2. Sensor information
Read the information on transmitter's unit, upper-lower limits, and the minimum span.
3. Device output
Table 8 Transmitter device output
Subject
Valid Value
Description
Process Variable/Range
Units
°C, °F, °R, K
Measurement unit of the transmitter
Lower Limit of PV Range
Support numeric input, lower limit expanding 10%
Lower limit of the process variable
Upper limit of PV Range
Support numeric input, upper limit expanding 10%
Upper limit of the process variable
Transfer Function
Linear, Square Root
Transfer function of the transmitter
Alarm State
Read only
Alarm state of the transmitter
Damping
Support numeric input, ≥0
Damping time
Poll Address
0~15
Poll address of the transmitter
Burst Mode
Enable/ disable
Burst mode state
Burst Command
1,2,3
Burst command depends on different
transmitters
3.2.3 Diagnosis and service
The diagnostic/service functions of the HART transmitter include current loop testing, primary variable zeroing, current
adjustment, range migration, and sensor trim.
Note: Current loop test, current adjustment, and other diagnostic adjustment functions that involve the use of current loop
Summary of Contents for 273E Series
Page 1: ...1 273Ex Handheld Pressure Calibrator...
Page 2: ......
Page 3: ......
Page 4: ...ADT273Ex Handheld Pressure Calibrator User Manual Version 2302V02 Additel Corporation...
Page 6: ......
Page 20: ...10 1 3 Basic Structure Figure 1 Basic Structure...
Page 41: ...31 Figure 16 Sensor trim Figure Figure 17 Process variable selection...
Page 56: ...46...