background image

COMPEX HD INSTRUCTIONS FOR USE

MADE IN THE USA

P

/N

 660202 

R

E

V. 

A 2021/10

!

ISO CERTIFIED

!
!

!

HEATED DISPENSER

!

Contents of package:

 (1) Compex HD unit, (1) wall plug-in USB power 

supply, (1) USB charging cord, (25) sample disposable barrier bags, (10) 

silicone sleeves, and (1) Instructions for Use.

PRECAUTIONS

 

  

 

    

Not Sterile – 

The silicone sleeves are patient contact parts and    

 

  

 are provided in a non-sterile condition.  They must be sterilized  

prior to initial patient contact and between patient applications. Refer to 

instructions for a) Cleaning and b) Sterilizing the silicone sleeve.

               

 

  

 The Compex unit can be disinfected to prevent cross  

 

 

 

  

 contamination.  The plastic housing should only be wiped with a    

 cloth and disinfectant.  Do not spray the plastic housing.  Refer to 

instructions for c) Disinfecting the Compex HD housing.

 

  

 DO NOT AUTOCLAVE THE COMPEX UNIT.

  Do not soak the 

 

  

 Compex unit in disinfectant solution. The Compex contains 

sensitive electronic components.

 

  

 Before heating and expressing composite, ensure that the compule 

 

  

 (PLT) is fully seated in the compule holder by inserting compule 

with even firm pressure.

 

 

  

 Attempting to express composite that has not been warmed to 

 

  

 operating temperature may result in damage to the unit.

 

  

 Battery Disposal: The battery pack contains rechargeable Lithium  

 

  

 Ion batteries (Li-Ion). Recycle or dispose of battery pack according 

 

  

 to national, state, and local regulations.

INTENDED USE

The Compex HD unit is intended to be used by a dentist or health care 

provider to warm and dispense dental composite materials contained in a 

compule or PLT (Pre-Loaded Tip) to 155 ̊F (68 ̊C).

COMPULES THAT FIT THE COMPEX DISPENSER

The Compex was designed to fit most compules on the market with a 

diameter of 6.32 mm to 6.55 mm. Please see our website for a 

comprehensive list. (www.addent.com)

          

           After continuous use of larger diameter compules, check to ensure     

           that smaller diameter compules have a secure fit in the compule 

holder prior to use.

PRODUCT SPECIFICATIONS

Compex Measurements: 

Unit Length: 5.75 inches (145 mm)

 

  

  

Unit Height: 5.2 inches (130 mm)

 

  

  

Unit Width: 1.2 inches (27 mm)

 

  

  

Unit Weight: 8.1 ounces (230 grams)    

                          

includes battery charger 

AC Supply Connection: 

100-240 Volts AC ± 10%, 50/60 Hz.

USB Power Supply: 

5 VDC, 2A

Battery Specification: 

Li-Ion single cell battery pack with safety circuit

 

  

  

Output: 3.6V, 3350mAh 

 

  

  

Type: Lithium Ion. 

 

  

  

Performance: Approximately 100 applications    

 

  

  

starting with 100% charge.

 

  

  

Battery Pack life: 500 full recharge cycles,  

 

 

  

  

approximately 5 years worth of use.

 

  

  

Safety: Batteries are protected from over  

 

 

  

  

charging, over discharging and short circuiting    

 

  

  

by means of a built-in safety circuit.

Charging Circuit Performance:   Fully discharged battery pack charges
 

  

  

to 100% in approximately 4.5 hours.

 

  

  

Smart charging circuit provides correct

 

  

  

voltage and current characteristics for

 

  

  

charging battery pack: Constant current

 

  

  

then constant voltage, 3.6 Volts,1.2A, @ 20°C.

Safety:     

This device complies with ANSI/AAMI/IEC  

 

 

  

  

60601-1, “Medical Electrical Equipment – Part    

 

  

  

1: General Requirements for Basic Safety and    

 

  

  

Essential Performance (Third Edition, 

 

  

  

Amend ment 1 2012).”

EMC:     

This device complies with ANSI/AAMI/IEC  

 

 

  

  

60601-1-2, “Medical Electrical Equipment,  

 

 

  

  

Part 1-2: General Requirements for Basic  

 

 

  

  

Safety and Essential Performance –    

 

 

  

  

Collateral Standard: Electromagnetic   

 

 

  

  

Disturbances – Requirements and Tests, 4th    

 

  

  

Ed. (2014).”

WARRANTY 

    

AdDent, Inc. hereby warranties that for a period of one year from date of 

purchase; this instrument shall be free from defects in materials and 

workmanship and will perform satisfactorily under normal use.

ACCESSORIES AND REPLACEMENT PARTS 

    

P/N 131031   Disposable Barrier Bags – box of 150

P/N 131032   Autoclavable Silicone Sleeves – box of 10

P/N 130002   Medical grade power supply

REPORTING OBLIGATION

Any event such as death or serious injury that has occurred to the patient, 

user or other persons in relation to use of this device should be reported to 

AdDent, Inc. and the competent authority of the Member State in which the 

user and/or patient is established.

PATENTS

This product is covered by one or more of the following patents with other 

patents pending: 6,312,254 • 6,616,448 • 7,097,452 • 10,172,690 • 

10,548,688 • EP1190681

Patents Insured By IPISC

MANUFACTURER

 

 

     

 

 

 

    AdDent, Inc. 

 

  

 43 Miry Brook Road, Danbury CT 06810 USA

 

  

 Phone: 203-778-0200

 

  

 Fax: 203-792-2275

 

    Email: [email protected]

 

    Website: www.addent.com

Reviews: