8250A Optical Power Meter Operation Manual
3.5.2 Calling the Preset Wavelength
3-10
3.5.2
Calling the Preset Wavelength
The four preset wavelengths in the wavelength setting can be called and set.
An example, in which a preset wavelength of 405 nm in No.1 is called, is shown below. See Figure 3-6.
The preset wavelength from No.0 to No.3 is called sequentially by pressing the
[
RECALL]
key.
If the wavelength, which is a value outside the wavelength range of the current connected sensor, is called,
“n FAIL” (n is the wavelength number) is displayed and the wavelength is not set.
Figure 3-6 Calling the Preset Wavelength
Key operation
Display
1.
[
]
1. 000
0780
The current set wavelength and wavelength sensi-
tivity coefficient are displayed.
(These values depend on the sensor.)
2.
[
RECALL]
0 SEt
0780
This display indicates that the preset wavelength
of 780 nm in No.0 is called correctly.
1. 000
0780
The display does not change because the same
wavelength as the previous wavelength is called.
3.
[
RECALL]
1 SEt
0405
This display indicates that the preset wavelength
of 405 nm in No.1 is called correctly.
0.220
0405
The new wavelength sensitivity coefficient is dis-
played.
(The value depends on the sensor.)
4.
[
]
Normal display
Setting complete.
•
If the wavelength, which is a value outside the wavelength range of the current connected
sensor, is called, “n FAIL” (n is the wavelength number) is displayed and the wavelength is
not set.
•
If the wavelength, which is a value inside the wavelength range of the current connected
sensor, is called, “n SEt” (n is the wavelength number) is displayed, the wavelength is set,
and a new wavelength and wavelength sensitivity coefficient are displayed.
Normal display
Determining
the setting
Preset wavelength
Example
Preset wavelengths are set in order from 0.
Wavelength sensitivity coefficient
Wavelength
Moving the cursor and changing the value