8250A Optical Power Meter Operation Manual
6.1.2 How to Test
6-4
NOTE:
If the measured current is without the accuracy in this test, perform the calibration according to 7,
“CALIBRATION” or request ADC CORPORATION to calibrate or repair it.
200
A
ZERO
0 A (Open)
70.0 nA
70.0 nA
FULL
190
A
355 nA
2,920 nA
2 mA
ZERO
0 A (Open)
0.70
A
0.70
A
FULL
1.9 mA
3.55
A
29.20
A
20 mA
ZERO
0 A (Open)
7.00
A
7.00
A
FULL
19 mA
35.5
A
292
A
50 mA
ZERO
0 A (Open)
70.0
A
70.0
A
FULL
50 mA
145
A
820
A
Table 6-3 Current Range Accuracy Specifications (2 of 2)
RANGE
Measurement
point
Setting value
Specification
823* series path
821* series path