8230E Optical Power Meter Operation Manual
A-3
A.2.1
Causes of the Generated Calibration Errors
Table A-3 Causes of the Generated Calibration Errors
No.
Cause Command
Displayed
err code
Status
Remarks
1
The IV offset calibration is not conver-
gent
XIVC
Err 10
ERR Bit7 ON
2
The difference between the VIG gener-
ation value, which is input when the
zero or full calibration is performed,
and the measurement value exceeds
50% of the full scale
The measurement value exceeds the full
scale
when calibrating
.
XPC
Err 10
ERR Bit7 ON
3
The full calibration is performed with-
out zero calibration
XPC
Err 15
ERR Bit13 ON
Command execution error
4
The VIG generation value is not within
the input range.
XPC
Err 15
ERR Bit13 ON
Command execution error