6247G/6247C DC Voltage Current Source/Monitor Operation Manual
2.2.4 Initializing Setting Conditions
2-34
2.2.4
Initializing Setting Conditions
The following procedure initializes the 6247G/6247C to the factory default settings.
However, the following items are not initialized.
•
Selected interface
•
GPIB address
•
Talk Only/Addressable
•
Header output
•
USB ID
•
RS-232 (baud rate, number of data bits, number of stop bits, parity)
•
Parameter load at power ON
Operation
Character display area
1.
Press
MENU
, and select
K)
PARAMETER
with
,
or
.
K) PARAMETER
2.
Press
to enter the Select level.
3.
Select
1) Param Load
with
,
or
.
1) Param Load
4.
Press
to enter the Input/Run level.
1) Param Load
Ent Load0
5.
Select
Ent Ld Dflt
with
,
or
.
1) Param Load
Ent Ld Dflt
6.
Press
ENTER
.
7.
When loading is complete, "
Done
" is displayed and
the menu reverts to the Select level.
1) Param Load
Done
1) Param Load
8.
Press
EXIT
to exit the menu.