6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5. TECHNICAL REFERENCES
5-1
5.
TECHNICAL REFERENCES
This chapter describes the detailed functions for more accurate measurement.
5.1
DUT Connection
5.1.1
Note for Output Terminals
Figure 5-1 below shows the internal wire connection of the 6247G/6247C.
The output terminals are cut off from the internal circuits by the Operate/Standby relays in Standby status.
Figure 5-1 Internal Wire Connection
HI SENSE
HI OUTPUT
LO SENSE
LO OUTPUT
OPR/SUS/STBY
IM
VS
IS
VM
A
V
2-Wire/4-Wire (STBY)