6243/44 DC Voltage Current Source/Monitor Operation Manual
3.3.7 MENU key (Setting Parameters)
3-13
Interlock In: Sets Standby when the input signal rises from Lo
to Hi.
If the signal is in the Hi state or open state, opera-
tions will not be executed even if the operate com-
mand is sent from a key or GPIB.
Operate Signal Out:
Outputs a Lo signal when the 6243/44 is in the
Operate status and a Hi signal when in the standby
status.
Complete/Busy
Selects the function of the
COMPLETE OUT BUSY IN/OUT
signal input on the rear panel.
Complete Out Front:
Outputs a negative pulse signal at the start of mea-
surement.
Complete Out End:
Outputs a negative pulse signal at the completion
of the measurement and completion of a period.
Complete Out HI:
Outputs a negative pulse signal when the compar-
ator calculation result is HI.
Complete Out GO:
Outputs a negative pulse signal when the compar-
ator calculation result is GO.
Complete Out LO:
Outputs a negative pulse signal when the compar-
ator calculation result is LO.
Busy In:
Becomes the busy input signal for synchronized
operation. When the input signal is Lo, measure-
ment or sweep step operations do not take place
until the signal becomes Hi.
Busy Out:
Becomes the busy output signal for synchronized
operation. Outputs a Lo level signal from the start
of the source delay until a period is completed and
measurement is completed.
Ext. Signal Width
Selects the output pulse width of the SYNC OUT and COM-
PLETE OUT signals on the rear panel.
20
s:
The output pulse width becomes approximately
20
s to 30
s.
100
s:
The output pulse width becomes approximately
100
s to 200
s.