6243/44 DC Voltage Current Source/Monitor Operation Manual
3.3.7 MENU key (Setting Parameters)
3-12
3.3.7.5
COMPARATOR (Comparison Operation Parameters)
Selecting
COMPARATOR
and pressing the key enables selection of parameters related to comparator
calculation.
Compare
Toggles comparator calculation between ON and OFF.
ON:
The following comparator calculations are per-
formed.
Upper Data
Measurement result; HI
Lower Data
Measurement result
Upper Data;
GO
Measurement result
Lower Data; LO
OFF:
Comparator calculation is set to OFF.
Upper Data
Enables the upper limit value setting for comparator calculation.
Lower Data
Enables the lower limit value setting for comparator calculation.
Result Buzzer
Selects the condition on which the buzzer sounds in accordance
with the results of comparator calculation.
HI:
The buzzer sounds when the result of the compar-
ator calculation is HI.
GO:
The buzzer sounds when the result of the compar-
ator calculation is GO.
LO:
The buzzer sounds when the result of the compar-
ator calculation is LO.
OFF:
The buzzer is OFF.
3.3.7.6
EXT.SIGNAL (External Control Signal Parameters)
Selecting
EXT.SIGNAL
and pressing the key enables selection of parameters related to the external
control signal input on the rear panel.
Operate Signal
Selects the function of the
INTERLOCK OPERATE IN/OUT
signal input on the rear panel.
Operate OFF In:
Sets Standby when the input signal rises from Lo
to Hi.
Operate ON/OFF In:
Sets Standby when the input signal rises from Lo
to Hi, and sets Operate when the signal falls from
Hi to Lo.