6240B DC Voltage Current Source/Monitor Operation Manual
5.2.16 Self Test
5-70
5.2.16
Self Test
The 6240B can self-test internal operations by turning on the power, executing the remote command, or
manual operation.
1. For more information on the self-test items and output results, see Table 5-19.
Table 5-19 Self-Test Items (1/2)
Error code
Description
Execution method
Message
TER register (*1)
Power ON
*TST?
Key
operation
Register
Data
1
001
ROM check SUM
ROM Chk SUM
-
-
2
002
Display section ROM/RAM
Panel Comm
3
-
LCA data
(Continuous Buzzer ON)
4
004
RAM read/write
RAM Rd/Wt
5
005
Analog section communication
Analog Comm
6
012
CAL data SUM
CAL data SUM
a
2
7
013
Parameter SUM
Param SUM
4
8
501
Calibration data lost
CAL dt Lost
16
9
502
Saved data lost
Save dt Lost
32
10
503
Saved parameter data lost
Para dt Lost
64
11
130
SCI communication error
No resp SCI
128
12
101
AD operation IR1/IR2 ratio
AD Ratio 1-2
b
1
13
102
AD operation IR2/IR3 ratio
AD Ratio 2-3
2
14
103
AD operation IR3/IR4 ratio
AD Ratio 3-4
4
15
104
AD operation IR4/IR5 ratio
AD Ratio 4-5
8
16
111
Analog section SRT line test
ADRST Sig
16
17
112
Analog section TRIG line test
ADTRG Sig
32
18
151
AD operation ZERO X10
ADx10 Zero
64
19
152
AD operation ZERO X1
ADx1 Zero
128
20
201
VSVM 300 mV ZERO
VSVM 0.3V Z
c
1
21
202
VSVM 300 mV +FS
VSVM 0.3V +FS
2
22
203
VSVM 300 mV -FS
VSVM 0.3V -FS
4
23
204
VSVM 3 V ZERO
VSVM 3V Zero
8
24
205
VSVM 3 V +FS
VSVM 3V +FS
16
25
206
VSVM 3 V -FS
VSVM 3V -FS
32
32
207
VSVM 15 V ZERO
VSVM 15V Z
64
33
208
VSVM 15 V +FS
VSVM 15V +FS
128
34
209
VSVM 15 V -FS
VSVM 15V -FS
256
35
211
High Limit 300 mV +FS
HL 0.3V +FS
512
36
212
High Limit 300 mV -FS
HL 0.3V -FS
1024
37
213
High Limit 3 V +FS
HL 3V +FS
2048
38
214
High Limit 3 V -FS
HL 3V -FS
4096
43
215
High Limit 15 V +FS
HL 15V +FS
8192
44
216
High Limit 15 V -FS
HL 15V -FS
16384