ADC Diagnostix Coax Ophthalmoscope Instruction Manual Download Page 18

18

Conducted HF

interference

according to

IEC61000-4-6

Radiated HF

rerference cording

to C61000-4-3

3 Vrms

150 kHz to 80MHz

3 V/m

80 MHz to 2.5GHz

Not applicable

10 V/m

3 V/m

Immunity

Tests

IEC 60601

Test Level

Electromagnetic

Environment Guidelines

Guidelines and manufacturer's declaration - electromagnetic immunity

This 

Diagnostix

TM 

Instrument

model is intended for operation in the electromagnetic environment specified below.

The customer or the user of this 

Diagnostix

TM 

instrument 

should ensure that it is used in such an environment.

Compliance

Level

Portable and mobile radio
equipment should not be used
within a distance from the
Diagnostix

TM 

instrument, including

cables, that is less than the
recommended safety distance as
calculated by the equation that is
appropriate for the transmission
frequency.

Recommended safety distance:
d = 1.2

ì

P

d = 1.2

ì

P 80 MHz to 1000 MHz

d = 2.3

ì

P 1400 MHz to 2.5 GHz

Where P is the nominal power of
the transmitter in Watts (W) as
specified by the manufacturer of
the transmitter, and d is the
recommended safety distance in
meters (m).

The field strength of stationary
radio transmitters should be less
than the compliance level at all
frequencies as verified by an on-
site test

a

Interference is possible in the
vicinity of equipment marked with
the following symbol

Note 1:

At 80 MHz and 800 MHz, the higher value applies.

Note 2: 

These guidelines may not apply in all situations. The propagation of electromagnetic waves is influenced by reflection and absorption by

buildings, objects and people.

a.

The field strength of stationary transmitters, such as base stations of wireless telephones and mobile field radio services, amateur radio

stations, AM and FM radio and television transmitters cannot be precisely determined theoretically in advance. In order to determine the
electromagnetic environment due to stationary HF transmitters, an investigation of the location is advisable. If the field strength determined
at the location of the Diagnostix

TM 

Instrument exceeds the compliance level indicated above, then the Diagnostix

TM 

Instrument must be

monitored with regard to its normal operation at each place where it is used. If unusual performance characteristics are observed, additional
measures such as re-alignment of the Diagnostix

TM 

Instrument or its removal to another place may be necessary.

b.

In the frequency range of 150 kHz to 80 MHz, the field strength should be smaller than 3 V/m.

Summary of Contents for Diagnostix Coax Ophthalmoscope

Page 1: ...1 Questions Call ADC toll free 1 800 232 2670 Otoscope Coax Ophthalmoscope Dermascope Throat Illuminator Power Handles Instruction Manual ...

Page 2: ...ument Heads 9 6 Diagnostix Otoscope 9 10 7 Diagnostix Ophthalmoscope 10 12 8 Diagnostix Dermascope 12 13 9 Diagnostix Throat Illuminator 13 10 Replacing the Lamp 13 11 Spare Parts and Accessories 13 12 Cleaning and Disinfection 14 13 Technical Specifications 15 14 Maintenance 15 15 Electromagnetic Compatibility 15 19 16 Warranty 19 17 Quality Standards 20 18 How to Contact Us 20 ...

Page 3: ...3 DiagnostixTM Otoscope DiagnostixTM Coax Ophthalmoscope DiagnostixTM Dermascope DiagnostixTM Throat Illuminator DiagnostixTM PMV Otoscope DiagnostixTM Coax Plus Ophthalmoscope ...

Page 4: ...d accessories to ensure safe and functional use of this product 3 WARNINGS AND PRECAUTIONS You have purchased a high quality ADC DiagnostixTM instrument which has been manufactured to the highest global standards and is subjected to rigorous quality control Read these instructions for use carefully before putting the unit into operation and keep them in a safe place If you should have any question...

Page 5: ...rectedintothepatient seyebe limitedtotheminimumlevelwhichisnecessaryfordiagnosis Infants aphakesandpersons withdiseasedeyeswillbeatgreaterrisk Theriskmayalsobeincreasedifthepersonbeing examinedhashadanyexposurewiththesameinstrumentoranyotherophthalmicinstrument usingavisiblelightsourceduringtheprevious24hours Thiswillapplyparticularlyiftheeye hasbeenexposedtoretinalphotography Warning Otoscope MUS...

Page 6: ...echargeable Handle 5560 3 5v requires desk charger Plug In Rechargeable Handle 5460 3 5v 230v or 120v Note These handles are compatible with ADC Riester and Welch Allyn 3 5v instrument heads Welch Allyn is a registered trademark Riester is a registered trademark 4 3 Inserting and Removing Rechargeable Batteries Screw off the battery cap on the lower part of the handle Depending on which handle you...

Page 7: ...ossible light output we recommend always fitting high quality batteries as described in 4 3 If you suspect that liquid or moisture could have entered the handle it must not be charged under any circumstances This could lead to a life threatening electric shock especially in the case of plug in handles To extend the service life of the battery the battery should only be charged once the light inten...

Page 8: ...ent heads are equipped with a security feature Use enclosed hex key to lock head onto power handle To secure insert hex key into slot on head and tighten by turning clockwise Keep the hex key in a convenient place in case the head must be removed for cleaning or service NOTE Locking the head is recommended on wall mounted systems only 4 6 Adjusting the Light Intensity With the ADC AdtronicTM elect...

Page 9: ...el Lens for Magnification Standard Series The swivel lens 3x max is fixed to the device and can be swiveled 360 Fig 3 PMVSeries The focusing wheel allows you to adjust the focusing range The adjustment wheel moves up and down to focus the lens Fig 4 6 4 Insertion of External Instruments into the Ear Standard Series only If you wish to insert external instruments into the ear e g tweezers you have ...

Page 10: ...ection lens can be adjusted on the lens wheel The following correction lenses are available Coax 3 5v Ophthalmoscope Fig 6 Plus 1 10 12 15 20 40 Minus 1 10 15 20 25 30 35 Coax Plus 3 5v Ophthalmoscope Fig 7 Plus 1 45 in single steps Minus 1 44 in single steps The values can be read off in the illuminated field of view Plus values are displayed in green numbers minus values with red numbers 7 3 Ape...

Page 11: ... a dilated pupil and general examination Coax Coax Plus Karo Grid For topographic determination of retina changes Coax Plus Only Slit To help determine levels of tumors and lesions Coax Coax Plus Fixation star Measuring eccentric fixation or locating lesions Coax Coax Plus 7 4 Changing Filters Using the filter wheel the following filters can be switched for each aperture Coax Coax Plus Ophthalmosc...

Page 12: ...average service life 15 h Coax Plus 3 5 V ophthalmoscope 3 5v 29 mA average service life 100 000 h 8 DIAGNOSTIXTM DERMASCOPE 8 1 Purpose The Diagnostix Dermascope described in these Operating Instructions is produced for early identification of changes of skin pigmentation malignant melanomas 8 2 Focusing Focus the magnifying glass by rotating the eyepiece ring Fig 10 8 3 Contact Plates Two contac...

Page 13: ...inator LED 3 5 V 28 mA mean life span 100 000h 10 REPLACING THE LAMP All Instrument Heads Remove the instrument head off the battery handle The lamp is located at the base of the instrument head Pull the lamp out of the instrument head with thumb and forefinger or a suitable tool Insert a new lamp CAUTION The pin on the ophthalmoscope lamp must be inserted into the guide groove on the instrument h...

Page 14: ...ection The instrument heads and handles can be cleaned externally with a moist cloth until visually clean Wipe disinfection as specified by the disinfectant manufacturer Only disinfectants with proven efficacy should be used taking into account the national requirements After disinfection wipe the instrument down with a moist cloth to remove possible disinfectant residues The components that come ...

Page 15: ...ation equipment can influence medical electrical equipment This ME device is intended for operation in an electromagnetic environment as specified below The user of the device should ensure that it is operated in such an environment The ME device must not be used directly next to or arranged in a stack with other devices If the device has to be operated near to or in a stacked arrangement with oth...

Page 16: ...he DiagnostixTM instrument employs HF energy solely for an internal function Its HF emission is therefore very low and it is unlikely that neighboring electronic devices will be affected by interference HF emissions according to CISPR 11 Class B Harmonics emissions according to EC61000 3 2 Not applicable The DiagnostixTM instrument is intended for use in all facilities including living quarters an...

Page 17: ...d output lines 3 A m 3 A m Not applicable The quality of the supply voltage should correspond to that of a typical business or hospital environment Surges IEC61000 4 5 1 kV voltage phase to phase 2 kV voltage phase to earth Not applicable The quality of the supply voltage should correspond to that of a typical business or hospital environment Voltage dips short time interruptions and fluctuations ...

Page 18: ...radio transmitters should be less than the compliance level at all frequencies as verified by an on site testa Interference is possible in the vicinity of equipment marked with the following symbol Note 1 At 80 MHz and 800 MHz the higher value applies Note2 Theseguidelinesmaynotapplyinallsituations Thepropagationofelectromagneticwavesisinfluencedby reflection and absorption by buildings objects an...

Page 19: ...2 5GHz For transmitters whose nominal power is not indicated in the table above the distance can be determined using the equation belonging to the respective column where P is the nominal power of the transmitter in Watts W as specified by the manufacturer of the transmitter Note 1 At 80 MHz and 1400 MHz the distance for the higher frequency range applies Note2 Theseguidelinesmaynotapplyinallsitua...

Page 20: ...QUALITY STANDARDS Device standard This device is manufactured to meet the European and United States standards for ISO 62471 2008 ISO 15004 ISO 10942 ISO 15004 1 and ISO 15004 2 Electromagnetic compatibility Device fulfills the stipulations of the International standard IEC60601 1 2 18 HOW TO CONTACT US To register your product and obtain further detailed user information about our products and se...

Reviews: