7. SPECULA COMPARTMENT
The specula compartment may be used for spare specula and lamps.
8. WALL MOUNTING
The ADC Diagnostix Desk Charging Base may
be mounted on the wall.
To Mount:
Hold wall rail in place on the wall
where intended for mounting. Level and mark
hole positions with a pen. Drill 5 mm (3/16”)
wide holes. Insert the supplied mollys in the
holes. Position spacer holes over the mollys
and affix rail over the spacer and secure to
the wall with the screws supplied (fig. 1).
Slide unit bracket guide slots of the ADC
®
Diagnostix
TM
Desk Charging Base plate onto the wall rail. When fully
inserted, the bottom cover tab will engage and lock in
the recess of the wall rail (fig. 2). Position the power
cord either to the right or left behind the base to
accommodate electrical outlet (fig. 3).
To remove:
Press the bottom tab lock,
which allows the unit to
slide from the rail (fig. 4).
Spacer
Rail
(fig. 1)
(fig. 2)
(fig. 3)
(fig. 4)
9
8
9. LED INDICATORS
10. CLEANING AND DISINFECTION
Prior to cleaning or disinfection of the unit disconnect mains cable
from socket.
Cleaning and/or Disinfection
The ADC Diagnostix Desk Charging Base may be cleaned externally with
a damp cloth.
In addition, the following disinfectants may be used for external disinfection:
Aldehyde (formaldehyde, glutaraldehyde, aldehyde derivatives), surfactants
or alcohols.
After disinfection, wipe the instruments with a damp cloth to
remove any remaining disinfectant.
Never place the charging station in liquids.
Use a soft lint free cloth or cotton
swabs to prevent damage to charging base.
Sterilization
According to the Test Centre for Medical Devices in Tu
̈
bingen, Germany,
sterilization is only prescribed in the case of operative procedures. There is
no necessity for sterilization of this charging base.
Yellow LED on
Charging in progress
LEDs switching on repeatedly,
brief permanent switch-on,
followed by off
Mains connection, self-test
Green LED on
Yellow LED flashing
Rechargeable battery fully charged, unit
changing to maintenance charge
Rechargeable battery fully discharged or
faulty. Unit trying to regenerate rechargeable
battery by a pulsating charging current.
Should no charging cycle be indicated
after approx. 16 h, the rechargeable
battery is faulty or must be replaced