ADC Advantage Fingertip 2200 Directions For Use Manual Download Page 6

Troubleshooting

Problem

SpO2 or pulse rate cannot be
shown normally

SpO2 or pulse rateis shown unsta-
bly

The Oximeter cannot be powered
on

Display suddenly powers off. 

“Err 3” or “Err 4” 

“Err 7” 

Symbol Definitions

The following symbols are associated with your Advantage™ 2200

The equipment type is BF

                           

Refer to user manual before use

                           

Hemoglobin saturation

                           

Heart Rate (BPM) Beats per Minute

Low Power Indication

Manufacturing Date

Conforms to EU Standards

Not for continous monitoring

Limited Warranty

American Diagnostic Corporation (ADC®) warrants its products against defects in materi-

als and workmanship under normal use and service as follows: 

1.     Warranty service extends to the original retail purchaser only and commences with the

date of delivery.

2.     Your Pulse Oximeter is warranted for one (1) year from date of purchase (all parts).

       

What is Covered:

Replacement of parts, and labor.

       

What is Not Covered:

Transportation charges to and from ADC®. Damages caused by

abuse, misuse, accident, or negligence. Incidental, special, or consequential damages.

Possible Reason

1.  Finger is not inserted correctly.
2.  Oxyhemoglobin value is too low to
be measured 

1.  Finger might not be inserted deep
enough.
2.  Excessive patient movement.

1.  No battery or low battery power.
2.  Battery not installed correctly.
3.  The Oximeter may be damaged.

1.  The Oximeter is automatically pow-
ered off, when no signal was detected
after 8 seconds.
2.  Low battery strength.

1.  Low Power.
2.  Mechanical device error near sen-
sor

1.  Low Power
2.  Oximeter is damaged

Solution

1.       Re-insert/re-position finger 
2.       Ensure that sensor is clean and not
blocked
3.       Try taking a reading on a different
finger or warm your finger by rubbing.

1.  Re-insert/re-position finger.
2.  Sit calmly and retry.

1.  Replace battery.
2.  Remove and reinstall battery.
3.  Contact customer service center.

1.  Replace the batteries
2.  Re-insert/re-position finger

1.  Replace the batteries.
2.  Return to Service Center

1.  Change batteries.
2.  Return to Service Center

PR BPM

Reviews: