Important! 

The small white arrow (Artery Mark) on the cuff

must lie exactly over the artery which runs down the inner side of
the arm.

d) Tighten the cuff by pulling the end and close the cuff by affix-
ing the hook and loop closure.

e) There should be little free space between the arm and the cuff. 
You should be able to fit 2 fingers between your arm and the cuff.
Clothing must not restrict the arm. Any piece of clothing which does,
must be removed. Cuffs that don’t fit properly result in false measure-
ment  values. Measure your arm circumference if you are not sure of
proper fit. Other size cuffs are available (Page 9).

f) Lay your arm on a table (palm upward) so the cuff is at the same
height as your heart. Make sure the tube is not kinked.

g) Remain seated quietly for at least two minutes before you begin the measurement.

Comment:

If it is not possible to fit the cuff to your left arm, it can also be placed on your 

right arm. However, all measurements should be made using the same arm. Comparable 
blood pressure measurements always require the same conditions (Relax for several 
minutes before taking a measurement).

5.4. Measuring procedure

After the cuff has been appropriately positioned on the arm and connected  
to the blood pressure monitor,  the measurement can begin:

a) Press the ON/OFF/START button.  The  pump begins to inflate the cuff. 
In the display, the increasing cuff pressure is continually displayed.

b)    After automatically reaching an individual pressure, the pump stops
and the pressure slowly falls. The cuff pressure is displayed during the
measurement.

c) When the device has detected your pulse, the heart symbol in the 
display begins to blink and an audible pulse tone sounds.

d) When the measurement has been concluded, a long beep tone sounds.
The measured systolic and diastolic blood pressure values, as well as the
pulse are now displayed.

e)The appearance of this symbol          signifies that an irregular heartbeat

was detected. This indicator is only a caution. It is important that you 
be relaxed, remain still and do not talk during measurements. 

 

NOTE:

We recommend contacting your physician if you see this

             indicator frequently.

f) The measurement results are displayed until you switch the device
off. If no button is pressed for 1 minute, the device switches off auto-
matically.

13

Pumping

Pressure

Measuring

Measurement

complete

Systolic

Diastolic

Pulse

Summary of Contents for Advantage 6021

Page 1: ...INSTRUMENT MUST BE USED ACCORDING TO INSTRUCTIONS TO ENSURE ACCURATE READINGS Questions Call ADC toll free at 1 800 232 2670 Digital Blood Pressure Monitor Advanced Blood Pressure Monitors Models 602...

Page 2: ...ghout the world where accuracy and dependability are critical ADC professional diagnostic products are the instruments of choice Now you too can enjoy the benefits of ADC engineering and quality in th...

Page 3: ...e Standard or ADC Averaging Mode Technology 10 4 5 Select the user 10 4 6 ADC Averaging Mode Technology 10 5 Measurement Procedure 11 5 1 Before measurement 11 5 2 Common sources of error 11 5 3 Fitti...

Page 4: ...is NOT intended to replace regular medical checkups It is recommended that your physician review your procedure for using this device Blood pressure readings obtained by this device should be verifie...

Page 5: ...of California to cause cancer birth defects or other reproductive harm To avoid any possibility of accidental strangulation keep this unit away from children and do not drape tubing around your neck...

Page 6: ...ry high if your diastolic pressure is above 90 mmHg and or your systolic blood pressure is over 160 mmHg while at rest In this case please consult your physician immediately Long term values at this l...

Page 7: ...in the vessel walls Arte riosclerosis A deficient supply of blood to important organs heart brain mus cles can result from arteriosclerosis Furthermore the heart will become structurally damaged with...

Page 8: ...smoking completely Drink only moderate amounts of alcohol Restrict your caffeine consumption coffee tea chocolate D Your physical condition After a medical examination and with your doctor s approval...

Page 9: ...ed above 9 Each Model has its own specific cuff Model Item Number Description 6021 850 6013 D bar Nylon Cuff For arm circumference 22 32 cm or 8 7 12 6 6022 850 6022 Wide Range Soft Cuff For arm circu...

Page 10: ...ADC Averaging Mode Technology 6022 6023 models only This instrument enables you to select either standard standard single measurement or ADC Averaging Mode Technology automatic triple measure ment To...

Page 11: ...ysittinginanarm chairinaquietatmosphereforabouttenminutesbeforetakingameasurement Removeanygarmentthatfitscloselytoyourupperarm Alwaysmeasureonthesamearm normallyleft Alwayscomparemeasurementstakenatt...

Page 12: ...e the cuff wrap it around your arm and press the hook and loop closure together d There should be little free space between your arm and the cuff You should be able to fit 2 fingers between your arm a...

Page 13: ...before taking a measurement 5 4 Measuring procedure After the cuff has been appropriately positioned on the arm and connected to the blood pressure monitor the measurement can begin a Press the ON OF...

Page 14: ...automatically scroll through the three meas urements used in the measurement 5 5 Irregular Heartbeat Detector This symbol indicates that certain pulse irregularities were detected during the measureme...

Page 15: ...our measurement If your measurement has only one dot your measurement is in the green zone or Normal according to NIH standards If your measurement has two or three dots it is in the yellow zone or Pr...

Page 16: ...ord FULL will be displayed to signal the oldest value is being deleted before displaying the current reading All other Models 6022 6023 The new measured value will overwrite the oldest value without a...

Page 17: ...emory button allows you to set the alarm hour To confirm press the Date Time button c The minute display will now flash The minutes can be set using the Memory button To confirm press the Date Time bu...

Page 18: ...te and time and possibly also set alarm times must be reset the year number therefore flashes automatically after the batteries are replaced d To set date and time follow the procedure described in Se...

Page 19: ...ons 6022 6023 only This unit can be used in connection with your personal computer PC running the ADC Advanced PC Link Software Your PC will allow a capacity of monitoring 80 patients each with 1000 r...

Page 20: ...between systolic and diastolic is excessive Measure again carefully following proper cuff proce dures and ensure measurement under quiet conditions Error No Possible cause s Solutions Err 1 The tube m...

Page 21: ...read the following information and the points listed under Common sources of error Section 5 2 Repeat the measurement Blood pressure changes constantly The observed readings may accurately reflect yo...

Page 22: ...t be disposed of in ac cordance with the locally applicable regulations not with do mestic waste 8 1 Accuracy test Sensitive measuring devices must be checked for accuracy from time to time We recomme...

Page 23: ...the same algorithm as ADC B H S graded AA model number BP3BT0 A US patents 6 485 429 Irregular Heartbeat Feature 6 447 457 ADC Averaging Mode 11 Technical Specifications Weight 471 5 g with batteries...

Page 24: ...55 Commerce Drive Hauppauge NY 11788 Inspected in the U S A Made in China tel 631 273 9600 1 800 232 2670 fax 631 273 9659 www adctoday com email info adctoday com IB p n 93 6023 00 rev 3 Printed in C...

Reviews: