background image

6

Size:                                           

Weight:                                       

Display:                                      

Battery:                                       

Battery Life:                                

Measurement Time:

                                                  

Measurement Range:                 

Accuracy:                                    

Measurement Mode:                   

Audible Alarm:                            

Auto Turn Off:                             

Memory Feature:                        

Operation Environment:              

Storage Environment:                 

Expected Service Life:                

Standard:

                                                  

*30 second response time achieved in water bath test.

5”L x 3/4”W x 3/8”H (12.7cm L x 1.905cm W x .9525cm H)

0.388 oz (11 grams

LCD, Liquid Crystal Display)

1.55v (LR41) Alkaline

Estimated 4,500 measurements

Oral 30* Seconds / Rectal 25-35 Seconds / 
Axillary 90 Seconds

90.0°F to 109.9°F / 32.0°C to 42.9°C

+/- 0.2°F/0.1°C between 96.0°F to 107.0°F / 34.0°C to 42.0°C 
at a room temperature of 64.4°F to 82.4°F / 18°C to 28°C

+/- 0.4°F/0.2°C between 90.0°F to 96.0°F/ 32.0°C to 34°C 
and 107.0°F to 109.9°F/ 42.0°C to 42.9°C.

Real measurement with peak-hold method

1 beep at start-up. Continuous beep at measurement 
completion

In approximately 8 ±1 minutes

Stores last temperature taken

10 °C to  40 °C /50 °F  to 104 °F,  humidity of 15% to 95%
(noncondensing)

-13°F to 140°F (–25°C to 60°C), humidity of 15 % to 95 % 
(noncondensing)

5 Years

IEC60601-1-11     

Specifications