6

Er.2:

If the temperature did not register because of poor contact,

patient movement or poor placement, the device will beep 5 times

and display “Er.2”

Remedy:

Turn the unit OFF, wait 1 minute, turn it ON again and retake tempera-

ture using proper technique.

Er.3:

If the unit malfunctioned, the device will beep 5 times and

display “Er.3”

Remedy:

Do not use the thermometer – Call 1-800-232-2670 for

service.

Low Battery:

Battery icon is on continuously.

Remedy:

The battery has enough remaining life for approxi-

mately 10 more measurements

Drained Battery:

Battery icon appears as unit is turned on and

the device turns itself OFF.

Remedy:

Replace battery.

Standards:

    Complies with EN 12470-3 and ASTEM E 1112-00

How to care for this product:

•  Wipe the tip and probe with a cotton swab moistened with 70% isopropyl alco-

hol before & after each use. After cleaning, wait at least 2 minutes before taking a

measurement for the unit to re-adjust to room temperature.

•  Use only a soft dry cloth to clean the body of the thermometer (if necessary).

Never use an abrasive cleanser or submerge the thermometer in water or other liq-

uids.

Summary of Contents for Adtemp 427 Temple Touch

Page 1: ...AdtempTM 427 TempleTouchTM Thermometer USE CARE MAINTENANCE...

Page 2: ...treme heat as they may explode WARNING It is recommended to remove the batteries from the thermometer when it is not used for a long period Do not dispose of batteries in fire recharge put in backward...

Page 3: ...s for the unit to readjust to room temperature 7 If the thermometer has been stored at a cold temperature allow the unit to warm naturally to room temperature before using 8 Performance may be degrade...

Page 4: ...re will be dis played for about 1 minute 7 Press and release the Multi Function button to turn the thermometer off or it will shut off automatically after 1 minute If you need to take another reading...

Page 5: ...f the measured temperature is above 107 6 F 42 C the de vice will beep and display HI Remedy Turn the unit OFF wait 1 minute and then ON again Re take temperature using proper technique If still reads...

Page 6: ...battery has enough remaining life for approxi mately 10 more measurements Drained Battery Battery icon appears as unit is turned on and the device turns itself OFF Remedy Replace battery Standards Co...

Page 7: ...o 42 C 60 8 F 104 F 16 C 40 C Humidity 15 95 non condensing 4 F to 122 F 20 C 50 C Humidity 15 95 non condensing Celsius or Fahrenheit Complies with ASTM E 1112 00 standard re quirements TN LCD 0 63 X...

Page 8: ...sulting from improper handling or use Specific legal rights stipulated in your state may guide this warranty If you have a legitimate claim under this warranty send item s postage paid to Service Cent...