10.4.3
Dynamic (Animal) Weighing
The balance can be set to weigh animals or unstable/moving items. This is commonly
referred to as ‘Dynamic’ or ‘Animal’ weighing mode. The balance will collect the weight
over a period of time to arrive at an average value and display the value until the operator
resets the balance. The actual weighing process can begin either automatically when the
weight is placed on the pan, or when initiated by the operator. The weighing unit can be
selected as normal using the [Unit] and [Setup] keys, before starting the dynamic
weighing process.
Steps:
Press [Mode] and then the [Up] or [Down] key to cycle through available modes. When
the “
Dynamic
” symbol is displayed, press [Setup] to enter dynamic weighing mode. “
rvn
rvn
rvn
rvn
” will
now be displayed on the screen.
Press [Up] or [Down] to select “
rvn
rvn
rvn
rvn
” for starting the dynamic weighing, or “
SEtvP
SEtvP
SEtvP
SEtvP
” to set
up the balance for dynamic weighing (see section 10.4.3.3 on Dynamic Weighing Setup
Parameters).
During dynamic weighing, if the [Print] key is pressed, the balance will display “
PAvSEd
PAvSEd
PAvSEd
PAvSEd
” for
1 second, then show the current average weight with the “
Dynamic
” symbol flashing.
To resume, press [Print] again or if you do not wish to continue then pressing [Mode] will
display “
StOP
StOP
StOP
StOP
” for one second and then show the final value. The value will be locked until
the user releases it.
10.4.3.1 MANUAL MODE
When the balance is in the “
MmANvAL
MmANvAL
MmANvAL
MmANvAL
” mode:–
If [Setup] is pressed when “
rvN
rvN
rvN
rvN
” is selected, balance will display “
StArt
StArt
StArt
StArt
”.
Place the item on the pan and press [Setup] again.
After the pre-configured delay and test time have elapsed (see section 10.4.3.3 on
Dynamic Weighing Setup Parameters), the “
Hold
” symbol and the result will be displayed.
Remove the item from the pan. Press [Mode] to go back to “
rUN
rUN
rUN
rUN
” to weigh another item, or
[
0/T
] to return to normal weighing.
10.4.3.2 AUTO MODE
When the balance is in the “
AvTO
AvTO
AvTO
AvTO
” mode:–
If [Setup] is pressed when “
rVN
rVN
rVN
rVN
” is selected, the balance will display “
LOAd P
LOAd P
LOAd P
LOAd P
”.
Place the item on the pan. The animal weighing test will begin automatically.
After the pre-configured delay and test time have elapsed (see section 10.4.3.3 on
Dynamic Weighing Setup Parameters), the “
HOLD
” symbol and the result will be displayed.
Remove the item from the pan. Press [Mode] to go back to “
rUN
rUN
rUN
rUN
” to weigh another item, or
[
0/T
] to return to normal weighing.