background image

 

 

3.0  LOCATING THE SCALE 

 
 

  Place the scale on a stable surface. A soft and uneven floor (e.g. carpet, 

linoleum) could cause an erroneous weight reading. 

 

  The ambient temperature of the scale should be kept as constant as possible. Do 

not place the scale where it may be subjected to draughts or heater and direct 
sunlight. It is advisable not to place the scale near windows, air conditioning vents 
or radiators. 

 

  The scale should be placed in a location where it is not likely to be damaged by 

accidentally dropping heavy weights or moisture on to the scale. The weighing 
area should be kept clean.  

 

  Clean the scale with a damp cloth whenever necessary. The bottom of the scale 

should be cleaned with a dry cloth only. Never use chemical or abrasive products. 

 

  It is a precision scale and all care should be taken to protect the scale from 

vibration, humidity, temperature variations and dust. 

 
 
 

4.0  SETTING UP THE SCALE 

 
 

  The MIW-20 comes with an internal alkaline battery supplied as standard with the 

scale. 
 

  Install the battery by removing the battery cover located at the bottom of the main 

unit. After connecting the battery, it is to be inserted in the socket provided at the 
back of the battery cover and then place the cover back in its original position. 

 

  There is no assembly necessary, simply press the 

[ON/OFF/TARE]

 key. 

 

 
 
 
 
 

© Adam Equipment Company 2004 

4

 

Summary of Contents for MIW-20

Page 1: ...Adam Equipment MIW 20 BABY SCALE ADAM EQUIPMENT CO LTD p n 4295 Rev B March 2004 Adam Equipment Company 2004 ...

Page 2: ... Adam Equipment Company 2004 ...

Page 3: ...ICATIONS 3 3 0 LOCATING THE SCALE 4 4 0 SETTING UP THE SCALE 4 5 0 KEY DESCRIPTIONS 5 6 0 OPERATIONS 5 7 0 SELECTION OF UNITS 6 8 0 HOLD FUNCTION 7 9 0 AUTOMATIC POWER SWITCH OFF 7 10 0 CALIBRATION 8 11 0 ERROR MESSAGES 8 Adam Equipment Company 2004 ...

Page 4: ...ight in kilograms or pounds The scale is fast accurate and simple to use It operates using an internal alkaline battery PP3 9V supplied as standard with the scale It has HOLD function to lock the display until the operator releases it Automatic turn off feature after 30 seconds is available to maximise the battery life Filters are optimised for weighing active babies Adam Equipment Company 2004 2 ...

Page 5: ...TIONS Specifications MIW 20 Maximum Capacity max 20 Kg 44 4 lbs Readability d 10 g 1 2 oz Tare range T 20 kg 44 4 lbs Battery One PP3 9V Operating temperature 5 C 35 C Net weight 2 15 kg 4 73 lbs Adam Equipment Company 2004 3 ...

Page 6: ...should be kept clean Clean the scale with a damp cloth whenever necessary The bottom of the scale should be cleaned with a dry cloth only Never use chemical or abrasive products It is a precision scale and all care should be taken to protect the scale from vibration humidity temperature variations and dust 4 0 SETTING UP THE SCALE The MIW 20 comes with an internal alkaline battery supplied as stan...

Page 7: ... the scale UNIT Sets the Units HOLD Holds the average display value 6 0 OPERATIONS To start weighing turn the scale on using the ON OFF TARE key It displays 0000 first If the chosen unit of weight is Kilogram next it displays 0 00 with a triangle at the top left of the display If the display shows 0 0 with a triangle at the bottom left of the display the unit is in pounds and ounces Adam Equipment...

Page 8: ...le first and pressing the ON OFF TARE key to zero the display Then the baby to be weighed can be placed on the cradle The display will show the weight of the baby without the added weight of the blankets or pillows 7 0 SELECTION OF UNITS Turn the scale on by pressing ON OFF TARE Press the UNIT key to toggle between Kilograms and Pounds If the chosen unit of weight is Kilogram next it displays 0 00...

Page 9: ... time or when the display is zeroed or tared with no weight on the scale 9 0 AUTOMATIC POWER SWITCH OFF The scale is operated using one PP3 9V alkaline battery in the internal battery chamber The typical battery life is 12 hours The scale includes the feature to turn the power off after 30 seconds if it is not being used This helps to maximise the battery life If the battery is low on charge LO wi...

Page 10: ...ht onto the scale Display will show a span count between 1500 to 3500 Press the ON OFF TARE key again Display will show CAL After 3 seconds the display will show 20 0 Remove the weight and keep the ON OFF TARE key pressed for 2 seconds The calbration is complete and the scale will be turned off 11 0 ERROR MESSAGES Error Messages Possible causes Suggestions LO If the batteries are low on charge LO ...

Page 11: ...ction against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the use...

Page 12: ...ctronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk Adam Equipment Inc 26 Commerce Drive Danbury CT 06810 USA Phone 1 203 790 4774 Fax 1 203 792 3406 e mail sales adamequipment com Adam Equipment S A Pty Ltd P O Box 1422 Kempton Park 1620 Johannesburg...

Reviews: