6.6
ADJUST
THE
VALUE
OF
THE
INTERNAL
MASS
The
internal
mass
value
stored
in
memory
can
be
adjusted
to
more
closely
match
the
value
of
the
user’s
external
mass.
1)
To
set
this
parameter
press
the
[Mode]
key
during
self
‐
checking
test
when
turning
on
the
balance.
2)
After
a
few
seconds,
the
display
will
show
the
first
function
F1 UNT
.
3)
Keep
pressing
the
[Mode]
key
until
F6 CA
is
displayed.
4)
Press
the
[Tare]
key
to
view
the
current
settings.
5)
The
display
will
show
the
current
value
with
the
first
digit
flashing.
To
change
the
value
press
[%]
to
shift
the
position,
press
[Mode]
to
increase
the
value
and
press
[Print]
to
decrease
the
value.
The
internal
calibration
weight
should
only
change
between
95.000
and
105.000
grams
or
495.00
and
505.00
grams
(depending
upon
the
model).
6)
Press
[Tare]
to
store
a
particular
setting.
The
display
will
return
to
F6 CA
7)
When
F6 CA
is
displayed
press
the
[Print]
key
to
return
to
weighing
or
press
[Mode]
to
go
to
the
next
function.
8)
Pressing
[Mode]
will
show
TECH
.
This
function
is
a
manufacturing
function
for
qualified
technicians
only
and
you
can
press
[Mode]
again
to
scroll
through
parameter
options,
or
press
[Print]
to
return
to
normal
weighing.
Changing
the
value
of
the
internal
weight
will
affect
the
calibration
when
using
the
internal
calibration.
To
check
if
your
internal
weight
is
correct
you
should
only
use
high
quality
weights
that
are
accurate
to
the
readability
of
the
balance.
19
|
P a g e
©
Adam
Equipment
Company
2008