22
© Adam Equipment Company 2012
7.5
AUTO
POWER
OFF,
F5
F5
F5
F5
OF
OF
OF
OFFFFF
The auto power off function helps conserve power by automatically turning the
balance off when it is not in use. The Auto power off time can be set by the user in
minutes, or to not be in operation (OFF).
1)
Switch on the balance then press and hold the
[Unit/Clr]
key during the self-
checking test of the display.
2)
After a few seconds, the display will show the first function
F1 UNT
.
3)
Keep pressing the
[Air/Liquid]
key until
F5 OFF
is displayed.
4)
Press the
[Tare]
key to view the current setting.
5)
Press
[Air/Liquid]
to change the setting to (Off, 5, 10, 20 or 30 minutes).
6)
Press
[Tare]
to confirm and move to the next setting.
7)
When
F5 oFF
is displayed press the
[Tare]
key to enter the parameter, the
[Unit/Clr]
key to return to weighing, or press
[Air/Liquid]
to go to the next
function.
7.6
INTERNAL
OR
EXTERNAL
CALIBRATION,
F6
F6
F6
F6
IIIICE
CE
CE
CE
You can select to use the internal mass or an external mass to calibrate the balance.
1)
Switch on the balance then press and hold the
[Unit/Clr]
key during the self-
checking test of the display.
2)
After a few seconds, the display will show the first function
F1 UNT
.
3)
Keep pressing the
[Air/Liquid]
key until
F6 ICE
is displayed.
4)
Press the
[Tare]
key to view the current setting.
5)
To change the setting press the
[Air/Liquid]
key
to select either
Int
CAL
(internal calibration) or
Etl
CAL
(external calibration).
6)
Press
[Tare]
to confirm and move to the next setting.
7)
When
F7 CA
is displayed press the
[Tare]
key to enter the parameter, the
[Unit/Clr]
key to return to weighing, or press
[Air/Liquid]
to go to the next
function.