4) The balance will initially show
SP 10
asking for a sample size of 10 parts.
Press
[Mode]
to cycle through the options: 10, 20, 50, 100, 200 and back to 10
to select the sample you have placed on the balance.
5) Press
[Smpl]
again once you have selected the sample. The display will show
you the number. As more parts are added the display will show the total
number of parts (
PCS
– will be shown in the alpha‐numeric display).
6) By pressing the
[Mode]
key you can view the unit weight (
W/P
), total weight
(
g
) or the count (
PCS
). The total and the unit weight are shown in the current
weighing units.
7) Press
[Smpl]
to return to normal weighing.
5.3
ACCUMULATION
There are two types of accumulations – automatic and manual.
Automatic Accumulation
Manual Accumulation
When the balance is set for automatic
accumulation,
see
Parameters
6.3
Setting the Printer Parameters /
Accumulation
, the weight will be added
to the memory when the balance
becomes stable.
When the balance is set to manual
accumulation,
see
Parameters
6.3
Setting the Printer Parameters /
Accumulation
, the weight displayed will
only be stored in memory once the
[Print]
key is pressed and the weight is
stable.
1) Place the weight on the balance. If automatic then the balance will
automatically accumulate the weight. If set to manual then you will need to
press the
[Print]
key.
2) The display will show
ACC 1
followed by the total value in the memory for 2
seconds before returning to displaying the weight of the item on the balance.
3) The weight will be transmitted to a printer or PC.
4) Remove the weight, allowing the balance to return to zero.
5) Put a second weight on. If set to automatic it will accumulate the weight
automatically. If in manual mode, press
[Print]
, the display will show
ACC 2