15.2.8 F8 –ZERO MODE
To enter this parameter, press the
[Tare]
key when
“F8 ZEm”
is shown.
Select the Zero mode desired. In all but special cases Zero Mode 1 is used. The other 2
zero modes are for unique locations in the world and effect the +/- range of the zero.
Press the
[Func]
key to change the value.
Press
[Tare]
to accept the displayed value.
Press
[Zero]
to return to weighing.
15.2.9 F9 –filter settings
This parameter allows the level of filtering to be defined, the slower the filter setting the
longer it will take to reach a stable reading, but the more accurate the results will be. In
contrast the faster the filter setting, the quicker it will reach a stable reading, but the results
may not be as accurate.
To enter this parameter, press the
[Tare]
key when
“F9 FL”
is shown.
The display will show the current filter setting which can be
SLoWES, SLoWEr, FAStEr
or
FAStES
.
Press the
[Func]
key to change the value.
Press
[Tare]
to accept the displayed value.
The display will show the level of filtering, which can be set at 1, 2, 3, 4, 5 or 6.
Press the
[Func]
key to change the value.
Press
[Tare]
to accept the displayed value.
Press
[Zero]
to return to weighing.
15.2.10 F10 –CALIBRATION COUNT
This parameter allows the calibration and parameter counting function to be active.
To enter this parameter, press the
[Tare]
key when
“F10 Cn”
is shown.
The display will show if the Calibration Counting Mode is set to
on
or
oFF.
If
On
, the
Calibration count and Parameter count will be seen at power on as described in section
6.0. If set to
oFF
the only method that can be used for access to the calibration or
parameters is to place the jumper on pins 1-2 of ADJ, See section 15.2.