© Adam Equipment Company 2011
12.0 SERVICE
PARAMETERS
12.1 USING
“0000”
TO
ENTER
THE
CALIBRATION
PARAMETER
Press the
[Tare]
key during the display counting when turned on. The
display will ask for a code number,
“Pin “
on the Weight Display.
Enter the number
0000
when
“Pin “
is displayed and then press
[Tare]
.
If another PIN number has been previously stored enter that number
instead.
The displays will show the first parameters, called
“F1” “CAL”
.
To select another parameter press the
[U.Wt./Units]
key to advance
through the parameters.
Press the
[Tare]
key to enter a parameter.
To exit a parameter, press the
[Zero]
key.
The Weight window will show the parameter number and the Unit Price
and Count windows will show the word describing the function.
When a parameter is entered by pressing the
[Tare]
key, the displays
will guide you through the parameter selected and the options
available.
The parameters available are:
“F1”
“
CAL”
To enter the Calibration
“F2” “Pin”
To set a new Pin number for access
“F3” “Unit”
Enable Weighing Units