7.0
TROUBLE
SHOOTING
7.1
ERROR
MESSAGES
If
an
error
message
is
shown,
repeat
the
step
that
caused
the
message.
If
the
error
message
is
still
shown
then
contact
your
dealer
for
support.
ERROR
CODE
DESCRIPTION
POSSIBLE
CAUSES
Solutions
ADO
A/D
Value
too
high.
Weight
on
the
pan
when
turning
on.
Improper
calibration
of
the
balance.
Damaged
load
cell.
Damaged
Electronics.
Remove
any
weight
from
the
stainless
steel
pan.
Recalibrate.
ADL
A/D
Value
too
low.
Pan
is
not
installed
when
turning
on.
Calibration
not
correct.
Damaged
load
cell.
Install
pan
and
power
back
on.
Recalibrate.
-----
A/D
count
is
not
correct
when
turning
the
balance
on.
Load
cell
damaged.
Electronics
damaged.
Remove
any
weight
from
the
stainless
steel
pan
and
try
rebooting.
Try
calibrating
the
balance.
Unit-E
Unit
error.
Low
battery
indicator.
Batteries
may
be
flat.
Charge
the
battery.
FAIL
Calibration
failure.
Incorrect
mass
used
to
calibrate.
The
user
calibration
is
not
within
5%
of
factory
calibration.
Possible
damaged
to
loadcell.
Try
to
recalibrate,
check
that
the
weight
used
to
calibrate
matches
that
of
the
value
of
the
balance.
Unstable
–
machine
cannot
get
a
stable
reading.
Possible
damage
to
the
Loadcell. Make
sure
the
balance
is
on
a
flat
surface
and
away
from
vibration.
No
Power
when
turning
on.
Battery
may
be
flat.
AC
adapter
may
not
be
working.
Change
the
batteries.
Or
use
the
AC
adapter
15
|
P a g e
©
Adam
Equipment
Company
2008