© Adam Equipment Company 2007
15
8.7
F7
-A/D COUNT
To enter this parameter, press the
[Tare]
key when “
F7 Cnt
” is shown.
This parameter allows you to view the A/D counts from the internal A/D converter. This can
be an aid to service.
Press the
[Tare]
key to return to the PARAMETER menu.
Typical value at zero is 30,000-90,000 (approx.)
Typical value at full capacity is 500,000 (approx.)
8.8
F8
–ZERO
This parameter allows the normal method of setting zero.
To enter this parameter, press the
[Tare]
key when “
F8 ZE M
” is shown.
The display will show the current settings, “
modE 1
” or “
modE 2
”.
Press the
[Func]
key to scroll through the options.
Press
[Tare]
to accept the displayed value and return to the parameter or press the
[Zero]
key to return to the parameter without changing the value.
Then press
[Func]
to advance to the next parameter.