Adam Equipment AE 301 MD-100 Manual Download Page 9

© Adam Equipment Company November 2011 

7

 

 

 

 

 

3.5  WEIGHING 

 

  Place the sample on the scale platform. The weight will be 

displayed in the selected Unit.  

 

  Wait until the stable indicator is displayed for accurate 

weighing. 

 

 

3.6  CHECK WEIGHING 

 

Check-weighing  is  a  procedure  to  sound  an  alarm  when  the  weight  of 
the  sample  on  the  scale  meets  or  exceeds  the  pre-set  Low  and  High 
values stored in the memory. 

 

To set the High and Low values, follow the procedure given below- 

 

 

Press  the  [High/Low] key.  The  display  will  show  “

000.00  HI

” 

with  the  first  digit  flashing  for  setting  the  High  value.  To  set 
your limit, use the [Unit] key to advance the flashing digit, the 
[Tare] key to decrease the value of the flashing digit and the 
[Zero] key to increase the value of the flashing digit .

 

 

  The  display  will  now  show  “

000.00  LOW

”  with  the  first  digit 

flashing  for  setting  the  Low  value.  To  set  your  limit,  use  the 
[Unit]  key  to  advance  the  flashing  digit,  the  [Tare]  key  to 
decrease the value of the flashing digit and the [Zero] key to 
increase the value of the flashing digit . Press [Mode] again 
to  confirm  the  value  and  the  display  will  return  to  normal 
weighing mode. 

 

  Once  the  check  weighing  Limits  are  set  and  the  alarm 

function is active, a beep will sound slowly when the sample 
weight is below the Low Limit. The beep will sound quickly if 
the weight is above the High Limit. If the sample weight is in 
between the Limits, the beep does not sound. 

Summary of Contents for AE 301 MD-100

Page 1: ... Adam Equipment Company November 2011 Adam Equipment AE 301 INDICATOR P N 700660194 Revision B November 2011 Software rev 2 4 ...

Page 2: ... Adam Equipment Company November 2011 ...

Page 3: ... 7 ALARM 8 3 7 1 Over load alarming 8 3 7 2 Low load alarming 8 3 7 3 Low battery alarming 8 4 0 USER PARAMETERS 8 4 1 BLT BACKLIGHT CONTROL 9 4 2 CHK CHECKWEIGHING CONTROL 9 4 3 FIL FILTERING CONTROL 9 4 4 SDT RS 232 CONTROL 10 4 5 PAR SETTING OF PARITY 10 4 6 BPS SELECTING BAUD RATE 10 5 0 RS 232 INTERFACE 11 6 0 TECHNICAL PARAMETERS 12 6 1 ZERO TRACKING 12 6 2 UNIT SETTING 12 6 3 CALIBRATION 13...

Page 4: ... Adam Equipment Company November 2011 2 ...

Page 5: ...interface as standard Selectable communication mode Operation from internal rechargeable battery or mains power 4 weighing units kg g ounce pound 2 0 SPECIFICATIONS Model AE 301 AE 301 MD 50 AE 301 MD 100 Load cell capacity 50 kg 100 kg Internal Counting resolution 250 000 Tare Range Full range Units of Measure 4 kg g lb oz Stabilisation Time 2 3 seconds Operating Temperature 0 C to 40 C Interface...

Page 6: ...sed during setting of the High and Low Limits for check weighing facility and for setting the alarm function Unit This key is used to change the weighing units while the indicator is in the weighing mode It is also used for advancing the digit while the indicator is in the setting mode Tare This key is used to deduct the weight of the container from the gross weight and thus to display the net wei...

Page 7: ... the AE 301 indicator to a new base it is necessary to connect the load cells properly Then set the parameters for the weighing system User should ensure the load cells are correct for the purpose and the signal from the load cell is correct Note that it may not be possible to measure the voltage signal from the load cell with all DVM Some DVM will measure the Excitation and signal voltage while o...

Page 8: ...er down 3 2 ZERO FUNCTION For accurate weighing zero the display by pressing the key Zero with nothing on the platform of the scale The zero indicator 0 will be on 3 3 TARE FUNCTION When the scale has a container on it press the Tare key The Tare indicator Net will be shown The display will show zero If it does not show zero press the Zero key To clear the tare remove the container and press the T...

Page 9: ... key to advance the flashing digit the Tare key to decrease the value of the flashing digit and the Zero key to increase the value of the flashing digit The display will now show 000 00 LOW with the first digit flashing for setting the Low value To set your limit use the Unit key to advance the flashing digit the Tare key to decrease the value of the flashing digit and the Zero key to increase the...

Page 10: ...ppears re charge the battery immediately to avoid any damage to the battery 4 0 USER PARAMETERS The following parameters can be set by the user by entering the function setting mode Press the Tare key while switching on the indicator The first parameter will be displayed along with the settings done previously The available parameters can be cycled through by pressing the Mode key Not all paramete...

Page 11: ...are 1200 2400 4800 and 9600 4 1 BLT BACKLIGHT CONTROL When BLT is displayed press the Unit key to change setting BLT 0FF Backlight is disabled BLT On Backlight is enabled Press the Mode key to store the setting and move on 4 2 CHK CHECK WEIGHING CONTROL When CHK is displayed press the Unit key to change setting CHK 0FF Check weighing is disabled CHK On Check weighing is enabled Press the Mode key ...

Page 12: ... 5 PAR SETTING OF PARITY When this function is displayed with the current setting press the Unit key to change to the other settings 8 n 1 8 data bits no parity default setting 7 E 1 7 data bits even parity 7 0 1 7 data bits odd parity Press the Mode key to store the desired setting and move to the next function 4 6 BPS SELECTING BAUD RATE When this function is displayed with the current setting p...

Page 13: ...weight printed Tare weight printed Net weight printed Blank line Blank line End of Transmission Input commands format The indicator can be used from a PC with the following commands The commands must be sent in upper case letters i e T not t and press the Enter button N Sends the net weight to the RS 232 interface G Sends the gross weight to the RS 232 interface T Sends the tare weight to the RS 2...

Page 14: ... other setting ZEO 0FF Zero tracking function is disabled ZEO On Zer0 tracking function is enabled Press the Mode key to store the desired setting and move to the next function 6 2 UNIT SETTING The user will be able to enable the weighing units to be used grams pounds or ounce Kilogram is the default weighing unit When g is displayed press the Unit key to change to the other setting g 0FF Gram uni...

Page 15: ...bration mass To select the desired mass press the Mode key The raw ADC count will be displayed along with CAI To set the zero calibration press the Mode key again The scale will display 0 LOD place the calibration mass selected onto the top pan the display will show the raw ADC counts of the selected calibration mass then press Mode again The display shows PASS and then returns to normal weighing ...

Page 16: ...ecure Obstruction under weighing platform Vibrations through the floor Temperature changed dramatically Power supply faulty Weight value incorrect Calibration error Recalibrate Unit calibrated with inaccurate weight Obstruction around platform Cannot use Full Capacity Overload stops hitting platform support or hitting bottom of load cell Parameters set incorrectly Load cell Damaged Not Linear Over...

Page 17: ...any other methods B Details of the unit purchased This part of information should always be available for any future correspondence We suggest you to fill in this form as soon as the unit is received and keep a print out in your record for ready reference Model name of the product Serial number of the unit Software revision number Displayed when power is first turned on Date of Purchase Name of th...

Page 18: ...iginal packaging and with correct documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This warranty does not cover equipment where defects or poor performance is due to misuse accidental damage exposure to radioactive or corrosive materials negligence faulty installation unauthorised modifications or attempted repair or failure to observe the requireme...

Page 19: ...ment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device Changes or modifications not expressly approved by Adam Equipment could void the user s authority t...

Page 20: ...ales Medical Scales Retail Scales for Price computing For a complete listing of all Adam products visit our website at www adamequipment com Copyright by Adam Equipment Co Ltd All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment Adam Equipment reserves the right to make changes to the technology fe...

Reviews: