COSMO
70
1
Info
CAUSE
REMEDY
204
Calculation overflow
Repeat procedure
252
Temperature too high
Cool down instrument
253
Temperature too low
Warm up instrument
255
Receiver signal too weak
Use target plate
256
Received signal too strong
Use target plate (grey side)
257
Wrong measurement
Use target plate (brown side)
258
Wrong initialization
Switch on – off the instrument