AcuTemp PX1L User Information Download Page 5

Rev. 6/26/07 

 

Page 5 of 9 

PERFORMANCE OF ACUTEMP PXC (-15

°

C)

x

AcuTemp PXC (-15

°

C)

packs are used with frozen vaccines or other frozen payloads.  

x

Used according to the instructions, the 

PX1L

 system will maintain vaccines/products at 

or below –15°C for more than 12 hours at 24°C (75°F). 

x

The hold time noted above requires the use of five 

PXC (-15

°

C)

coolant packs.  If three 

PXC

packs are used, the temperature maintenance time is reduced. 

x

As the ambient temperature increases, the hold time will decrease. 

HOW TO FREEZE ACUTEMP PXC (-15°C) PHASE CHANGE MATERIAL (PCM) 

x

Must be frozen in a freezer maintaining -21°C or below. 

x

PXC (-15°C)

 requires 48 hours at -21°C to freeze properly.  

HOW TO USE PXC (-15°) PHASE CHANGE MATERIAL (PCM) WITH THE 
ACUTEMP PX1L CARRYING CASE 

x

Avoid deforming the physical shape or the packaging of the 

PXC (-15

°

C)

.

x

Note that each 

PXC (-15

°

C)

will have a pull-tab available for ease of removal/use with 

the

PX1L

carry case.  When placing 

PXC

coolant packs in the 

PX1L

carrying case, be 

sure to keep the pull-tab “UP” so that the pack can be removed easily. 

x

Place one frozen 

PXC (-15

°

C)

on the bottom of the 

PX1L

carrying case.   This is the 

bottom PCM layer. 

x

Evenly distribute the payload on top of the first layer of PCM material with the payload 
height (H) not exceeding 2.0”.  (See Figure 1) 

x

Place the remaining four 

PXC (-15

°

C)

packs on top of the payload being careful to keep 

the removal tabs “UP”. 

x

In case of damage to 

PXC (-15

°

C)

, clean up and disposal may be handled with common 

soap and water. 

Summary of Contents for PX1L

Page 1: ...Temp PX1L Carrying Case and AcuTemp PXC Coolant Packs Formerly VaxiPac and VaxiSafe AcuTemp Thermal Systems 2900 Dryden Road Dayton Ohio 45439 U S A Tel 937 312 0114 Fax 937 312 1277 Email support acu...

Page 2: ...ibility of the shipper The unit must be packaged to avoid damage in transit AcuTemp is not responsible for any such damage AcuTemp reserves the right to change product specifications materials used in...

Page 3: ...the PXC 7 C x Note that each PXC 7 C will have a pull tab available for ease of removal use with the PX1L carry case When placing PXCs in the PX1L be sure to keep the pull tab UP so that the pack can...

Page 4: ...C requires 48 hours at 14 C to freeze completely x In order to properly freeze the PXC 10 C material in the AcuTemp AX27L place it in the bottom half of the AX27L unit with unit set to freeze HOW TO...

Page 5: ...C to freeze properly HOW TO USE PXC 15 PHASE CHANGE MATERIAL PCM WITH THE ACUTEMP PX1L CARRYING CASE x Avoid deforming the physical shape or the packaging of the PXC 15 C x Note that each PXC 15 C wi...

Page 6: ...28 C to freeze properly HOW TO USE PXC 20 PHASE CHANGE MATERIAL PCM WITH THE ACUTEMP PX1L CARRYING CASE x Avoid deforming the physical shape or the packaging of the PXC 20 C x Note that each PXC 20 C...

Page 7: ...Rev 6 26 07 Page 7 of 9 Figure 1...

Page 8: ...ying case Please be certain to keep the pull tab UP 3 Place your payload on top of the PXC pack making certain that there is enough space on top to add the remaining four PXC packs or 2 if you are onl...

Page 9: ...imum Weight five pieces 5 0 lbs 2 3 kg maximum PXC 10 C Dimensions maximum 1 1 28 mm H x 3 3 84 mm W x 9 3 236 mm L Weight one piece 1 4 lbs 630 gm maximum 1 2 lbs 530 gm minimum Weight five pieces 7...

Reviews: