Active forever Slendertone Manual Download Page 7

Technical Specifications

When the         symbol appears the stimulus is significantly
reduced.

The signal gradually increases to a peak intensity level at the start
of a contraction phase (i.e. ramp up) and gradually decreases to
nothing at start of a relaxation phase (i.e. ramp down).

Accessories

When ordering new pads, you should only use those carrying the

slendertone

brand. Any others may not be compatible with your

unit and could degrade the minimum safety levels. 

slendertone

Adhesive Pads, containing:

1 large adhesive pad

Type 709 or 706

2 small adhesive pads Type 710 or 707

slendertone

Belt

slendertone

Belt Extension

IIn

ntte

en

nd

de

ed

d  u

usse

e:: Muscle stimulator

W

Wa

avve

effo

orrm

m:: Symmetrical bi-phasic square waveform 

when measured into a resistive load.

Environmental Specifications

Operating:  Temperature Range: 0 to 35˚C

Humidity: 20 to 65 % RH

Storage: 

Temperature Range: 0 to 55˚C
Humidity: 10 to 90 % RH

Description of your unit’s symbols:

There are a number of technical markings on your unit. These can
be explained as follows:

Inside the battery compartment ‘+’ indicates positive polarity and
‘-’ indicates negative polarity. The batteries’ position is also
shown by 3 embossed outlines in the base of the compartment.

The unit and belt are manufactured for Bio-Medical Research
Ltd., Parkmore Business Park West, Galway, Ireland.

The unit requires three 1.5 volt 

(lr

03) DC batteries. DC is

indicated by the symbol:

The Output Frequency indicates the number of pulses per second
transmitted by the unit. This is measured in hertz ( ‘Hz’).

Output (RMSA) mean the max. output root mean square current
for each channel.

Output (RMSV) mean the max. output root mean square voltage
for each channel.

This symbol means ‘Attention, consult the accompanying
documents’.

SN stands for ‘serial number’.

Inside the box, on the back of the unit is the serial number
specific to this unit. The letter preceding the number indicates the
year of manufacture, where ‘A’ denotes 1995, ‘G’ denotes 2001, ‘H’
denotes 2002 etc.

The belt’s batch number is represented on the belt packaging, by
the number corresponding with the   LOT symbol.

Rated Outputs Voltage/Current:

P

Prro

od

du

ucctt  T

Ty

yp

pe

e:: 510 (female 

FLEX

)

Parameter

500Ω

1KΩ

1K5Ω

Output RMSV 

6.9V

12.9V

13.2V

Output RMSA

13.7mA

12.9mA

8.8mA

Output Frequency

40-70Hz.

40-70Hz.

40-70Hz.

DC Component: Approx.

0 C

0 C

0 C

Positive Pulse Width:

200-300 µs 200-300 µs 200-300 µs

Negative Pulse Width: 

200-300 µs 200-300 µs 200-300 µs

Interphase Interval:

100 µs

100 µs

100 µs 

P

Prro

od

du

ucctt  T

Ty

yp

pe

e:: 515 (male 

FLEX

)

Parameter

500Ω

1KΩ

1K5Ω

Output RMSV 

7.1V

13.4V

13.7V

Output RMSA

14.2mA

13.4mA

9.1mA

Output Frequency 

45-75Hz.

45-75Hz.

45-75Hz.

DC Component: Approx.

0 C

0 C

0 C

Positive Pulse Width:

200–300 µs 200–300 µs 200–300 µs

Negative Pulse Width:  200–300 µs 200–300 µs 200–300 µs
Interphase Interval:

100 µs

100 µs

100 µs

P

Prro

od

du

ucctt  T

Ty

yp

pe

e:: 514 (female 

FLEX MAX

)

Parameter

500Ω

1KΩ

1K5Ω

Output RMSV 

5.8V

9.3V

9.7V

Output RMSA

11.5mA

9.3mA

6.5mA

Output Frequency

50-95Hz.

50-95Hz.

50-95Hz.

DC Component: Approx.

0 C

0 C

0 C

Positive Pulse Width:

200-350 µs 200-350 µs 200-350 µs

Negative Pulse Width: 

200-350 µs 200-350 µs 200-350 µs

Interphase Interval: 

100 µs

100 µs

100 µs 

P

Prro

od

du

ucctt  T

Ty

yp

pe

e:: 517 (male 

FLEX MAX

)

Parameter

500Ω

1KΩ

1K5Ω

Output RMSV 

6.1V

9.7V

10.6V

Output RMSA

12.1mA

9.7mA

7.1mA

Output Frequency 

55-99Hz.

55-99Hz.

55-99Hz.

DC Component: Approx.

0 C

0 C

0 C

Positive Pulse Width:

200–350 µs 200–350 µs 200–350 µs

Negative Pulse Width:  200–350 µs 200–350 µs 200–350 µs
Interphase Interval:

100 µs

100 µs

100 µs

!

This symbol means type BF equipment.

This symbol on your 

slendertone

unit is to

indicate conformity to the requirements of the
Medical Device Directive (93/42/EEC). 0366 is
the number of the notified body (VDE).

Note:

In accordance with the laws within the member

states, safety testing on the device should be carried out
every two years.

Note: 

In compliance with German law, product safety

testing must be conducted every two years by an
authorised test service. 

0366

EU Only:

12

Problem

The display doesn’t come on &
there is no signal from the unit

The display is on but there’s 
no signal

The battery symbol is flashing

The contractions are very weak
even when the intensity is high

Unpleasant feeling beneath 
the pads

The       symbol has appeared on 
the display*

E002/E003/E004 has appeared 
on the display

Unintentional programme change
during Ab Crunch programmes 

Possible cause

The batteries are inserted
incorrectly

The batteries are dead

Low-quality batteries used

The unit has been paused

The batteries are low

The batteries are low

The pads are worn

Poor pad positioning

Pads not covering studs

Low-quality batteries used

The pads are worn

Pads not covering the studs

Too many consecutive sessions

Pads not in contact with skin

Pads not covering metal studs

The pads are worn

The pads are on the wrong 

The covers are still on the pads

Unit is not fully inserted in holder

Belt is loose or open on body

Error message, indicating a problem
with the unit 

Accidental pressing of keys 

Solution

Reposition the batteries as
described on page 5

Replace the batteries

Use only high-quality batteries

Press the on/off button briefly

Replace the batteries

Replace the batteries

Contact your local stockist for 
new 

slendertone

abdominal

replacement pads

See page 5 for the correct
positioning of the pads & belt

Reposition the pads

Use only high-quality batteries

Contact your local stockist for 
new 

slendertone

abdominal

replacement pads

Reposition the pads

You should leave at least six 
hours between sessions

Reposition the belt

Reposition the pads

Contact your local stockist for 
new 

slendertone

abdominal

replacement pads

Ensure the black surface of the
pads are facing you

Ensure the covers are removed 
from both sides of the pads

Reinsert the unit, see page 5 for
further details

Refasten belt, see page 5

for further details

See page 9 for further details

Activate intensity keylock. See 
page 9 for further details.

*  The        symbol always appears in conjunction with 
the left/right arrows. If the left arrow appears, there 
is a problem with the left pad, the right arrow indicates 

a problem with the right pad and if both arrows appear,
there is a problem with either the middle pad or all 
three pads.

Trouble-Shooting Guide

13

s l e n d e r t o n e   s u p p o r t   pa c k

s l e n d e r t o n e   s u p p o r t   pa c k

Summary of Contents for Slendertone

Page 1: ...ight 2003 Bio Medical Research Ltd All rights reserved slendertone slendertone flex slendertone flex max flexercise and c s i are trademarks of Bio Medical Research Ltd Irish Patent No S81372 UK Paten...

Page 2: ...njection short term or long term e g hormone treatment Please wait before using your slendertone Abdominal Training System until At least six weeks after the birth of your baby you must consult your d...

Page 3: ...programme see page 6 9 Increase the intensity to a level you find comfortable FIG f The unit will automatically increase the left or right intensity to ensure that one stays no more than 10 units abov...

Page 4: ...rcising immediately if you feel any pain Specific Do s and Don ts for SLENDERTONE Ab Crunch Programmes Never do abdominal crunches if you suffer from neck or lumbar pain Never exert yourself beyond yo...

Page 5: ...nly For FLEX users only Advanced Unit Features 1 Intensity Keylock Function If you find a comfortable intensity level press the upper and lower halves of the left intensity button to lock that intensi...

Page 6: ...ltation with your doctor is required to establish a rehabilitation programme with the SLENDERTONE unit which would safely provide improvement to the strength and tone of the abdominal muscles Caring f...

Page 7: ...th 200 350 s 200 350 s 200 350 s Negative Pulse Width 200 350 s 200 350 s 200 350 s Interphase Interval 100 s 100 s 100 s P Pr ro od du uc ct t T Ty yp pe e 517 male FLEX MAX Parameter 500 1K 1K5 Outp...

Reviews: