Only use recommended AC
adapter model. Unplug adapter to stop
using power.
* Do not wash the cuff in a washing machine or dishwasher!
* The service life of the cuff may vary by the frequency of washing, skin condition, and
storage state. The typical service life is 10000 times.
* It is recommended that the performance should be checked every 2 years and after
maintenance and repair, by retesting at least the requirements in limits of the error of the
cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).
* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according
to the local guidelines.
* Manufacturer will make available on request circuit diagrams, component part lists,
descriptions, calibration instructions,etc., to assist to service personnel in parts repair.
* The plug/adapter plug pins insulates the device from the main supply. Do not position the
device in a position where it is difficult to disconnect from the supply mains to safely
terminate operation of ME equipment.
* The operator shall not touch output of batteries /adapter and the patient simultaneously.
* Cleaning :Dust environment may affect the performance of the unit. Please use the soft
cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile
cleaners.
* The device doesn’t need to be calibrated within two years of reliable service.
* If you have any problems with this device, such as setting up, maintaining or using, please
contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself
in the event of malfunctions. The device must only be serviced, repaired and opened by
individuals at authorized sales/service centers.
* Please report to Transtek if any unexpected operation or events occur.
* Keep the unit out of reach of infants, young children or pets to avoid inhalation or
swallowing of small parts. It is dangerous or even fatal.
* Be careful to strangulation due to cables and hoses, particularly due to excessive length.
* At least 30 min required for ME equipment to warm from the minimum storage temperature
between uses until it is ready for intended use. At least 30 min required for ME equipment to
cool from the maximum storage temperature between uses until it is ready for intended use.
* This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS;
* Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this
equipment and should be kept at least adistance d away from the equipment. The distance
d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 and
Table 9 of IEC 60601-1-2:2014, as appropriate.
* Please use ACCESSORIES and detachable partes specified/ authorised by
MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the
user/patients.
* There is no luer lock connectors are used in the construction of tubing, there is a possibility
that they might be inadvertently connected to intravascular fluid systems, allowing air to be
pumped into a blood vessel.
* Please use the device under the environment which was provided in the user manual.
Otherwise, the performance and lifetime of the device will be impacted and reduced.
9
8
Monitor Components
List
Blood Pressure
3
Monitor
Instructions for Use
Upper-arm
Cuff (22cm-42cm)
4× AAA batteries
To p
rotect your device and obtain consistent readings, please use the
correct batteries and authorised power adapter
,
which complies with CE
safety standard
s
.
• Open the battery cover.
• Install batteries by matching the
correct polarity, as shown.
• Replace battery cover.
Installing and Replacing Batteries
CAUTION
CAUTION
Replace batteries when:
The
shows.
The display is dim.
The display does not light up.
Choice of Power Supply
AC adapter
Battery powered mode:
6V
DC 4×
AAA batteries
AC
adapter powered mode:
6V
(Optional not include
d
).
Do not use new and used batteries together.
Do not use different types of batteries together.
Do not dispose of batteries in fire. Batteries may explode or leak.
Remove batteries if the device is not being used for some time.
B
atteries are harmful to environment
, d
ispose
of correctly for your region
.
Remove old batteries from device and follow your local recycling guidelines.
UPPER-ARM
CUFF
AIR HOSE
AI
R
CONNECTOR
PLUG
LCD DISPLAY
MEM BUTTON
ST
ART/STOP BUTTON
BATTERY COMPARTMENT
SET BUTTON
DC POWER SOCKET
MODEL: TMB-1491-BS