4 Medical device description
4.1 First use
The Mini LED uses a Lithium-Ion battery. To guarantee optimum use, this battery must be fully charged before use but
must never be flat. To prepare your Mini LED, follow the steps below:
l
clean the rigid protection shield with an alcohol wipe;
l
clean the handpiece with an alcohol wipe;
l
plug the charging base into the mains;
l
check that the green indicator light on the base lights up twice and listen out for a beep sound;
l
install the battery on the handpiece;
l
place the Mini LED on its charging base;
l
the Mini LED is correctly installed when two beeps are heard and the green indicator light on the base starts
flashing;
l
leave the Mini LED to fully charge;
l
the Mini LED is charged when the green indicator light on the base stops flashing and remains on permanently;
l
remove the protection plug from the handpiece nozzle;
l
insert the sterilised optical guide into the handpiece nozzle;
l
you will hear a click when the optical guide is correctly inserted;
l
install the rigid protection shield;
l
wear safety goggles and protective gloves;
l
provide your patient with safety goggles.
4.2 Using the Mini LED
The Mini LED is normally placed on its base. To use it, remove it from its base.
Prior to the day of use, check that you have enough sterilised optical guides and check the power output as indicated in
the chapitre
Checking the lamp power level page 19
The patient and the practitioner must wear class II safety goggles when the Mini LED is in operation.
l
Remove the Mini LED from the base;
l
Press the on/off button.
l
Install the rigid protection shield;
l
Programme the Mini LED to select the required mode;
l
Position the end of the optical guide as close as possible to the surface of the material to be cured;
Do not allow the optical guide to touch the material to be cured.
l
Press the on/off button to trigger the selected mode. A beep sound confirms cycle initiation;
l
A second beep sound confirms cycle completion.
You can stop the cure cycle at any time by pressing the on/off button.
After three minutes of inactivity, the Mini LED switches to standby.
Depending on the cure material used, repeat the cure cycle as required.
4.3 Mini LED™
The Mini LED can only be used with the following accessories:
l
7.5 mm opalescent optical guide
l
7.5 mm amber optical guide
l
Mini LED titanium battery
l
Mini LED handpiece
l
Protection plug
l
Rigid protection shield
l
Charging base and mains adapter
The protection plug is designed to prevent any products from infiltrating the handpiece that may damage its
electronics, the connector or the LEDs. The protection plug must be installed when the handpiece is being cleaned.
Another transparent plastic protection plug protects the battery connectors.
User Manual • Mini LED™ • J02541 • V6 • (07) • 07/2016 • NG13EN010F - Page 11/30
Summary of Contents for Mini LED
Page 1: ...User Manual Mini LED ...
Page 6: ......
Page 8: ...Page 6 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Page 10: ...Page 8 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Page 18: ...Page 16 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Page 20: ...Page 18 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...
Page 30: ...Page 28 30 User Manual Mini LED J02541 V6 07 07 2016 NG13EN010F ...