Ai C20 RECIRCULATING CHILLER USER’S MANUAL
6
Model
C20-7
Electrical requirements
110V 60Hz 1-Phase 1500 watts
Environmental Operation
Conditions
•
Indoor use and altitude up to 2000m
•
Ambient Temperature
75 °F (24 °C)
•
Max. relative humidity 80% for temperatures up to 30ᵒC decreasing linearly to 50 % relative humidity at
40ᵒC
•
Mains supply volt fluctuations up to ±10 % of the nominal voltage
•
Transient overvoltage up to the levels of overvoltage category II (See Note 1).
•
Temporary overvoltage occurring on the mains supply.
•
Applicable pollution degree of the intended environment (pollution degree 2 in most cases (See Note 2).
Note 1: These levels of transient overvoltage are typical for equipment supplied from the building wiring.
Note 2: Manufacturers may specify more restricted environmental conditions for operation, nevertheless the
equipment must be safe within these normal environmental conditions.
Compressor
Make: SECOP (formerly known as Danfoss), Germany
Power: 115V 60Hz 1-PHASE
Number of compressor: one
Refrigerant: R404A (190g)
Cooling Capacity
1077 watts @ 0°C
535 watts @ -20°C
Heating Capacity
1100 watts
Recirculating pump
Pumping speed: 20 liters per minute, adjustable
Maximum head: 16 feet or 5 meters
Reservoir
Dimensions: 8 x 10 x 5.5" (WxDxH)
Material: Type 304 stainless steel, polished
Capacity: 7.3L or 1.9 gallons
Chilling fluid connection: 3/8" hose barb x 2 Chilling fluid
Drain port: 3/8" hose barb with on/off valve
Temperature controller
Type: Digital microprocessor PID controlled
Display: 7" TFT touch screen
Display unit: Degree Celsius °C
Temperature step: 0.1°C
Temperature accuracy: +/- 2°C
Temperature range: -20°C (-4°F) to 99°C (210°F)
Recommended coolant
10°C (50°F) or higher: Distilled water
Between 10°C (50°F) to -20°C (-4°F): 50% ethylene glycol and 50% distilled water
Built-in swivel casters
No
Unit dimensions (WxDxH)
12 x 15.5 x 26"
Shipping dimensions (WxDxH)
15 x 20 x 30"
Unit weight
68 lbs.
Shipping weight
100 lbs.
Safety
Compressor over-temp protection
Compressor power surge delay protection
Short circuit and ground protection
Over-current protection
Over-temperature protection
Upper/lower temperature alarm protection
Recommended for
Ai 2L, 5L or 10L short path distillation systems
Ai 2L SolventVap rotary evaporators