SmartReader Plus Data Logger
44
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
1.8
SmartReader Plus 5 Thermocouple Logger
The SmartReader Plus 5 Thermocouple Logger can monitor and record temperatures from J,
K, T, S or E type thermocouples.
Figure 5-1: SmartReader Plus 5
1.8.1
Description
The SmartReader Plus 5 has two external thermocouple temperature channels and one
internal thermistor channel.
For information on temperature ranges and resolutions applicable for the different types of
thermocouples you want to use, refer to the
Specifications
section.
1.8.2
Setup
The internal thermistor temperature channel for the SmartReader Plus 5 is used to simulate a
reference junction for the thermocouples - it must always be enabled.
Channel
Description
Equation
0
Internal Temperature
45
3
Thermocouple 1
see
SmartReader Plus 5 Specifications
5
Thermocouple 2
see
SmartReader Plus 5 Specifications
1.8.3
Use
1.
Connect a J, K, T, S or E type thermocouple to one or more channels. Make sure that
you observe polarity when making connections. To do this, connect a thermocouple's
negative wire (usually red) to the "-" terminal of the channel you want to use. Connect the
Summary of Contents for SmartReader Plus
Page 3: ...SmartReader Plus is a Registered Trademark of ACR Systems Inc TRADEMARK NOTICES ...
Page 8: ...SmartReader Plus Data Loggers IV Copyright 2010 ACR Systems Inc All Rights Reserved Index 105 ...
Page 96: ...SmartReader Plus Data Loggers 87 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 97: ...SmartReader Plus Data Logger 88 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 98: ...SmartReader Plus Data Loggers 89 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 116: ......