Use of Barriers – Intermediate Level Disinfection
The use of an all-purpose barrier film provides surface protection from cross-contamination resulting from a variety
of applications. This precaution should be used whenever dealing with non-intact skin or the chance of coming in
contact with bodily fluids. Barrier film is designed to cover any surface that may be touched during patient therapy,
in order to help prevent cross-contamination. Barrier film is for single-use only. The film is discarded after each
patient therapy. The procedure for use is as follows:
1.
Wash hands.
2.
Apply Intermediate Level Disinfection prior to barrier application.
3.
Select, tear or cut with clean scissors a length of barrier film to fit over the operator surfaces of the
Omnicycle
®
Elite unit.
4.
Prepare any items which may become in contact with the therapist during treatment, such as ultrasound gel,
pens, assessment tools, cart handles, etc.
5.
Set up the patient per guidelines for the therapy.
6.
Provide therapy as appropriate.
7.
Discard all disposables.
8.
With clean gloves, remove the plastic film from the unit and discard.
9.
Intermediate disinfect the Omnicycle
®
Elite unit prior to the next therapy application.