User manual
9
ACOWA INSTRUMENTS, Industrivej 10, 8305 Samsø - VAT-no: 36562919 – mail: [email protected] – Phone: +45 7221 7979
Installation
Power supply
PUMA must be connected to the supply voltage according to the specifications below.
Voltage supply
1x230V AC or 3x400V AC +10%/-20%
Frequency
50/60Hz
Input power consumption
0,004 -> 0,06A
Max supply fuse
16A
IK max
6kA
System grounding
TT
PUMA's communication module is equipped with a built-in power bank and
can send voltage failure alarms if the primary power supply stops working.
WARNING!
The unit must not be disassembled until the power bank is
switched off.
Physical measurements
For installation of PUMA, the following dimensions may be necessary.
Dimensions (W x H x D)
W=237mm x H=248mm x D=98mm
Weight
1350g
Wire connection
0.5 – 6 mm
2
Vibration (sinusoidal)
10-500Hz, 1G
Free fall drop
30cm
Enclosure class
IP65
Environment
Humidity
10% - 95% non-condensing air
Operating temperature
-20°C to +50°C
Storage temperature
-20°C to +60°C
Built-in power supply
PUMA has a built-in power supply which is intended for voltage supply of sensors and input and output
signals. Power supply V:
Output voltage
24V DC
Output current
Max 100mA
Tolerance
+/- 20%
Analog input
PUMA is equipped with one analog input 0...20 mA / 4…20 mA.
Number of analog mA inputs
1
Electrically insulated
No
Measuring range
0/4 – 20mA