16. Use the SET key to toggle the value to its desired setting
17. Press the
¿
key to store the value
18. The display should read
UP-
19. Press the SET
key to return to Operation mode (normal use)
User Parameter Specifications
User
Parameter
Definition
Range
UP-1
This controls the update speed on the display. Set
to 0 will make the display update the weight most
quickly; whereas a setting of 3 will update most
slowly.
0~3
UP-2
A setting of 1 corresponds to Average Weighing; a
setting of 0 sets the balance to Normal Weighing.
0~1
UP-3
A setting of 1 corresponds to disabling Auto-Zero
tracking; a setting of 0 sets the enables Auto-Zero
Tracking.
0~1
UP-4
A setting of 1 corresponds to Handshake RS-232
mode; a setting of 0 sets Stream RS-232 mode.
0~1
Average Weighing:
This mode is used whenever you need to weigh
items that may be in motion: liquids, livestock, etc.
Auto-Zero Tracking:
You will want to disable this feature when the
items that you weigh most often are very near zero.
Stream Mode:
When you use the RS-232 serial port in this mode, the
balance will automatically transmit the displayed data whenever the
balance becomes stable.
Handshake Mode:
When you use the RS-232 serial port in this mode,
the balance will only transmit the displayed data whenever the balance
is polled. This is usually done by connecting the balance to a PC.