WAN
Below is an example of Airodis setup for a WAN network. Ensure the IP address is set the same as on
the remote modem/router.
Figure 16 – WAN Network Set-Up (Airodis)
4.4
Digital/Analog Communication
The 25 Pin analog/digital port on the rear of the instrument sends and receives analog/digital signals
to other devices. These signals are commonly used to activate gas calibrators or for warning alarms.
4.4.1
Analog Outputs
The instrument is equipped with one to three analog outputs that can be set to provide either
voltage (0-5V, 0.25-5V, 0.5-5V, 0-10V) or current (0-20, 2-20, 4-20 mA). The analog outputs are tied
to specific instrument measurements, depending on the instrument type.
For 0-10V analog output operation, set the output type to current and move the jumpers (JP3) on the
rear panel PCA to 0-10V (refer to Figure 17).
Table 4 – Analog Outputs
Analyser
Output 1 Output 2
Output 3
S10
O
3
N/A
N/A
S30
CO
CO
2
[optional] N/A
S40
NO
NO
X
NO
2
S44
NO
NH
3
NO
2
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...