ACL 300B Operation Manual 

Rev. 6/9/10 KCB 

Page 5 of 5 

ACL MODEL #300B 

CALIBRATION INSTRUCTIONS 

Accuracy should be checked only at ½” spacing.  At greater distances, the viewing area of the electrode 
and therefore the area of the surface needed for calibration become prohibitively large. Additionally, 
stray electrostatic charges and nearby grounds become a source of error. 

EQUIPMENT REQUIRED 

1. Monroe Electronics, Inc. Model 241 ±3KV Reference Supply (or equivalent), capable of a 500V DC

and 3000V DC output at least 1% accurate.

2. A test fixture consisting of an electrically isolated metal plate at least 3” on a side (the larger, the

better) and a means of positioning this plate ½" from the electrode of the unit under test.

TEST PROCEDURE 

1. Connect the output of the Reference Supply to the test plate and the case (of the ACL 300B) to the

low side of the Reference Supply or ground.

2. Position the unit under test so that its electrode is ½" from the plate's center.
3. Mechanically zero the meter
4. Turn the unit on to LO.
5. Press the zero switch two or three times and the meter should zero.
6. Apply +450V DC to the test plate from the Reference Supply.
7. The meter of the unit under test should read b400 and +500 volts.
8. Release the applied voltage.
9. Repeat Step #5.
10. Repeat steps #6 & #7 using –450V DC.
11. Observe the meter for approximately 30 seconds and it should not drift more than one division

(50V). Twenty percent per minute drift maximum.

12. Release the applied voltage.
13. Set the unit under test to HI.
14. Repeat step #5.
15. Apply +3000V DC to the test plate.
16. The meter should read b2500 and +3500 volts.

NOTE:   Any accurate higher voltage may be used here between 2000 and 5000 volts and the meter 

should read the voltage applied ±10%. 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for ACL 300B

Page 1: ...10 KCB Page 1 of 5 ACL 300B Precision Static Locator OPERATION MANUAL Meter is warranted for one year from the date of purchase on parts and labor Calibration is recommended every twelve months www com information itm com 1 800 561 8187 ...

Page 2: ...arge relative to the sensitive electrode We refer to these relative sizes in our Field of View graph as Aspect Ratio OPERATION 1 Test meter before using following instructions on p 3 of this manual 2 Turn the meter on by pushing the red power range selector switch is into the LO or the HI position 3 Discharge static electricity from your body by touching a grounded metal object such as a water pip...

Page 3: ...n not in use by returning red switch to middle position 8 Remove the battery when storing the instrument for extended periods of time BATTERY REPLACEMENT To check battery press black BATT ZERO switch to left toward BATT side The needle should move to the right of the red battery line If it does not battery replacement is needed The battery should be replaced after forty hours of use or at least on...

Page 4: ...plastic 2 Rub balloon or plastic vigorously on arm 3 Using free hand press red switch on ACL 300B to the left in LO position 4 Pointing ACL 300B away from balloon or other charged objects press black BATT ZERO switch twice to the right to zero meter 5 Bring balloon toward meter s silver conductive sensor plate Needle should move freely to extreme right or left Caution DO NOT TOUCH BALLOON TO SENSO...

Page 5: ...ate and the case of the ACL 300B to the low side of the Reference Supply or ground 2 Position the unit under test so that its electrode is from the plate s center 3 Mechanically zero the meter 4 Turn the unit on to LO 5 Press the zero switch two or three times and the meter should zero 6 Apply 450V DC to the test plate from the Reference Supply 7 The meter of the unit under test should read betwee...

Reviews: