SHAVER SYSTEM
WARNING!
The application of the membrane suction-irrigation pump for small and closed body cavities (arthroscopy and
hysteroscopy) is strictly forbidden, because irrigation process may lead to uncontrolled and dangerous pressure rise.
4. DelIvery range
Check device and accessories immediately on receipt for completeness and transport damage. In case of damages please
inform your authorized dealer immediately.
If the unit has to be returned use only original box and attach the following information:
• Name and address
• Model No. / Serial No.
• Failure description
• Error message
ATTENTION!
Contaminated devices are not accepted for check or repair.
The unit consists of:
Pos.
Specification
Article no.
Qty.
1
Membrane suction-Irrigation pump
16-2083
1
2
Test report
n/a
1
3
User manual
n/a
1
4
Silicone suction-irrigation tubing set, autoclavable
16-2083-102/-103
1
5
Main cable Euro 2 m
16-2083-104
1
6
Fuses
16-2083-105
2
7
Secretion container with overflow valve, autoclavable
16-2083-101
1
8
Bottle holder (1 litre) + 2 screws for each
16-2083-106
2
9
Spike, autoclavable
16-2083-109
1
10
Hydrophobic sterile filter, sterile package, disposable
16-2083-108
2
5. SaFeTy InSTruCTIonS, DISpoSal, SaFeTy SIgnS & SymbolS
Please make yourself familiar with this unit according to description in the user manual.
Please note:
• The pump must only be used in endoscopic operations. Use of pump for other than mentioned procedures is not allowed.
Otherwise operator is responsible for any kind of risk.
• The type of irrigation liquid must be chosen in accordance with the selected surgical procedure.
• The device must only be used by trained specialists at medical institutions.
• For your and your patient‘s safety use only original accessories.
• Service and maintenance have to be carried out only by authorized persons.
• The device is not explosion-proof and must