5 Installing the device
This medical device is intended for use by a qualified surgeon for endoscopic applications.
No special training is required to install the device. Kindly refer to the instructions in this
manual.
5.1 Installation
Place the device on a stable surface and have on hand the various accessories required
for it to function;
Make sure the area is well ventilated.
Connect the power cord to the mains socket of the device [C3]; Check that the external filter
is properly in place on the CO2 input socket [C6]. This is necessary for the protection of the
internal circuits and will help prolong the life of the insufflator.
5.2 Connecting to a cylinder of medical CO²
Maximum allowed operating pressure (60 bars)
Never begin any surgical operation without having a CO2 replacement cylinder. Only
use medical-grade CO2, in compliance with European standards.
Place the CO2 cylinder in a vertical position, so that it is secure with the upper part
facing upwards, if possible on the same medical cart as the insufflator.
Never use the insufflator connected to a cylinder in a horizontal position or with the
upper part facing downwards.
Use the supplied open-end spanner to tighten the connector attached to the cylinder
outlet, and the connector attached to the external filter on the input port on the insufflator
[C6].
5.3 Connecting to a medical CO² central wall point
Manually connect the low-pressure hose to the wall connector .
Minimum allowed operating pressure: 3 bars.
Only use with the CO2 gas central wall system, in compliance with European standards.