Canister Full Therapy Interrupted Alarm
Medium Priority Alarm - This alarm screen appears when the canister is full and should be replaced. It is
indicated by a repeating audible tone.
Press
Audio Pause
to silence the audible tone.
The audible tone will come back on in two
minutes if the alarm condition has not been
fixed.
Therapy will stop during this alarm.
This alarm is fixed by changing the canister.
1. Use the graduated marks on the canister to check the
fluid level. A full canister is about 300 mL.
2. If canister is not full, press
Cancel
.
3. If the canister is full, change canister (see
Canister
Changes
section).
4. Press
Reset
to return to the Patient Mode
Home
screen.
5. Restart therapy by pressing
On / Off
.
WARNING: If the therapy unit will be off
for more than two hours, call your doctor
or nurse right away. Without power to the
therapy unit, your dressing will need to be
replaced.
To avoid a false alarm, keep the therapy
unit upright.
WARNING:
If Canister is full, replace and press ‘Reset’.
If not full, press ‘Cancel’. Press ‘?’ for more information.
Audio Pause
Reset
Cancel
Canister Full
Therapy Interrupted
Audio Pause
Cancel
Graduated Marks
Reset
On/Off