I. QC-100 OPERATING INSTRUCTIONS

1.  Remove and discard any used 

QuickCheck 

Vial  (QCV).    If  a  vial  is        

left attached to the QC-100 during storage the flow path may become        

saturated.  This is especially a concern if more than one gas type of 

QCV is being used.

2. 

Press the On/Off button to turn on the QC-100.  A green light         indicates 

normal operation.  A red light indicates a low battery condition.  If a red 

light occurs, replace the two ‘AA’ batteries with fresh, alkaline batteries 

and repeat step two. 

( Note:

  Do not use a metal object to remove the 

batteries as this can short out the electrical circuit.)

3. 

Select the QCV with the gas type desired.  

4. 

Squeeze the QCV firmly at the point where the label reads ‘

PInCh             

heRe TO bReAk InneR AmPOuLe

’.    This  allows  the  liquid  to         

absorb into the wicking material and start releasing the test gas.

5. 

Install the QCV into the QC-100,  To install insert the QCV snuggly         

into the opening on the right side of the instrument near the top.

6. 

Attach any tubing or calibration adapters if needed.  

Note: 

  A  three         

foot section of chemically resistant tubing is included with the QC-100.  

If inappropriate tubing or adapters are used the chemical may be lost 

in the sample system without reaching the sensor to be tested.

7. 

Place the end of the QC-100, tubing or calibration adapter at the sensor 

to be tested.

8. 

If an appropriate response is not achieved the output of gas can be 

varied  by  opening  (turning  counter-clockwise)  or  closing  (turning 

clockwise) the needle valve on the top side of the QC-100.

9. 

Repeat steps 7 and 8 for any other sensors to be tested, 

Note: 

Do not 

continue testing if the QCV is depleted.

Page 2

Summary of Contents for QC-100

Page 1: ...05USA Telephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 1100 00 AdvancedCalibrationDesigns Inc InstructionManual CalibrationGasBumpTester Adva...

Page 2: ...Instruction Manual QC 100 QuickCheck System TABLE OF CONTENTS I QC 100 Operating Instructions Page 2 II Accessory Items Parts List Page 4 III Specifications Page 5 910 1050 00...

Page 3: ...inner ampoule This allows the liquid to absorb into the wicking material and start releasing the test gas 5 Install the QCV into the QC 100 To install insert the QCV snuggly into the opening on the r...

Page 4: ...QCV 11 Note Once the inner ampoule is broken the QCV should be quickly inserted into the QC 100 Opened QCV deplete quickly without being installed Once testing is finished please remove and discard th...

Page 5: ...lable for the QC 50 P N Description 525 0001 12 NH3 Ammonia 12 Pack 525 0002 12 HCl Hydrogen Chloride 12 Pack 525 0003 12 C7H8 Toluene 12 Pack 525 0004 12 C6H14 Hexane 12 Pack 715 0405 0X Hose with Qu...

Page 6: ...ne year Dimensions QC 100 6 0 x 2 3 x 1 2 15 2 x 5 8 x 3 0 Dimensions QCV 0 4 x 1 3 1 1 x 3 4cm Weight QC 100 8oz 220gm Weight QCV 0 14oz 4gm Operating Temperature 5 50 C Relative Humidity intermitten...

Page 7: ...by Advanced Calibration Designs Inc 2024 W McMillan Street Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 email ACD goacd com website w...

Page 8: ...05USA Telephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 1100 00 AdvancedCalibrationDesigns Inc InstructionManual CalibrationGasBumpTester Adva...

Reviews: